MiPNet14.06 Instrumental O2 background

From Bioblast
Jump to: navigation, search

Oroboros Instruments

Products         Purchase         Support         O2k-Publications         Workshops & Events         O2k-Network         Feedback



MiPNet14.06 Instrumental O2 background

Publications in the MiPMap
O2k-Protocols
O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux.

» Bioblast pdf » Versions

Oroboros (2018-08-07) Mitochondr Physiol Network

Abstract: Fasching M, Gnaiger E (2018) O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6(07):1-14.

O2k-Protocols SOP: Correction for instrumental background oxygen flux is a standard in high-resolution respirometry, automatically performed by DatLab. Background measurements provide a test of instrument function. In the Oroboros O2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become highly significant and require compliance to standard operating procedures described in this chapter as part of the MitoFit Quality Control System. This is part 2 of O2k Quality Control.

» O2k Quality Control 1 »MiPNet06.03 POS-calibration-SOP«
» Product: O2k-FluoRespirometer, O2k-Catalogue
» MitoPedia: Instrumental background oxygen flux

Keywords: Instrumental O2 background

O2k-Network Lab: AT_Innsbruck_Oroboros


Labels: MiParea: Respiration, Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual 

DatLab, O2k-SOP 


Oroboros support

O2k-technical support

This information is part of O2k-technical support and open innovation.