Data recording interval

From Bioblast
Jump to navigation Jump to search

Header ecosystem 2020.jpg

OROBOROS support
        O2k-Manual         O2k-Videosupport         O2k-Procedures         MitoPedia: O2k hardware         MitoPedia: DatLab         MitoPedia: SUIT         MitoPedia: O2k-Open Support        


Data recording interval


Template NextGen-O2k.jpg
O2k-Open Support


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

The data recording interval is the time interval at which data is sampled with an instrument. In DatLab the data recording interval is set in the O2k control window. The system default value is 2 s. A lower data recording interval is selected for kinetic experiments, and when the volume-specific oxygen flux is high (>300 pmol O2·s-1·ml-1).
Technically, the O2k instrument (hardware) measures the sensor signal every 10ms (which is NOT the „data recording interval“). By the given data recording interval from DatLab (software) a discrete number of sensor signal points are taken to calculate an average value in the O2k (e.g. a data recording interval of 2 s can take 200 sensor signal points; a data recording interval of 0.5 s can take 50 sensor signal points). This average value is sent to DatLab and is recorded as a raw data point. However, there is a defined threshold: the O2k does not apply more than 200 sensor signal points to calculate the average for the raw data point. For example a data recording interval of 3 s could take 300 sensor signal points but only the 200 most recent sensor signal points are used for averaging.

Abbreviation: F7

Reference: MiPNet22.11 O2k-FluoRespirometer manual


MitoPedia O2k and high-resolution respirometry: DatLab