Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Template:Select fluorescence sensor

From Bioblast
Switching between different excitation wavelengths and filters is achieved by simply exchanging the Fluo-Sensors. Two types of optical sensors are supplied with different LEDs for fluorescence excitation, and the effective spectra of the LEDs are modified by filters.
Select the Fluo-Sensor and filter set from the table below in the section "Fluorescence dyes and settings". Each Fluo-Sensor is delivered with a mounted filter set:
  1. Fluorescence-Sensor Green and Smart Fluo-Sensor Green: 525 nm (dominant wavelength). The installed Filter-Cap is ready for H2O2 measurement with Amplex® UltraRed or for measurements of mt-membrane potential with TMRM.
  2. Fluorescence-Sensor Blue and Smart Fluo-Sensor Blue: 465 nm (dominant wavelength). The installed Filter-Cap is ready for measurements of mt-membrane potential with safranin or Rhodamine 123. A different filter is used for measurement with Magnesium Green™ or Calcium Green™.

PlayVideo.jpg Exchanging a filter set of the Smart Fluo-Sensor