Volume-Calibration Ring sV

From Bioblast
Jump to: navigation, search

Header ecosystem 2020.jpg

OROBOROS support
        MitoPedia: O2k-Open Support         MitoPedia: O2k hardware         MitoPedia: DatLab         MitoPedia: SUIT         O2k-Procedures         O2k-Manual         O2k-Videosupport        


Volume-Calibration Ring sV


Template NextGen-O2k.jpg
O2k-Open Support


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

Volume calibration ring A+B PVDF for Stopper.jpg

Volume-Calibration Ring sV (A or B), white PVDF, attached to PEEK O2k-Stopper sV.


Reference: Product details and purchase information



Template NextGen-O2k.jpg
O2k-Open Support


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Technical support

Volume calibration of the 0.5-mL O2k-chamber
  • Loosen the stopper fixation rings (A) and (B) with the Allen key.
  • Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
  • Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
Volume calibration ring A+B aluminum for Stopper.jpg


Keywords


Questions.jpg


Click to expand or collaps



MitoPedia O2k and high-resolution respirometry: O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry