Line 10: Line 10:
:::: '''As: Ascorbate sodium salt''' (Sodium L-ascorbate); Sigma A 7631, 100 g, store at RT; M = 198.1 g·mol<sup>-1</sup>.
:::: '''As: Ascorbate sodium salt''' (Sodium L-ascorbate); Sigma A 7631, 100 g, store at RT; M = 198.1 g·mol<sup>-1</sup>.


:::: <span style="color:#8B008B"> '''Caution:''' Light sensitive (store solution in a dark glass vial).</span>
:::: <span style="color:#8B008B"> '''Caution:''' Solutions of ascorbic acid are rapidly oxidized in air and in alkaline media; the compound should be protected from air and light. See 'ascorbid acid' on [https://pubchem.ncbi.nlm.nih.gov/compound/54670067 PubChem]</span>  
 
<!--
<!--
:::: '''Preparation of 800 mM stock solution''' (dissolved in H<sub>2</sub>O):  
:::: '''Preparation of 800 mM stock solution''' (dissolved in H<sub>2</sub>O):  
::::# Weigh 1.584 g of ascorbate sodium salt and dissolve in 10 mL H<sub>2</sub>O
::::# Weigh 1.584 g of ascorbate sodium salt and dissolve in 10 mL H<sub>2</sub>O
Line 26: Line 24:


-->
-->


:::: '''Preparation of 800 mM stock solution''' (dissolved in H<sub>2</sub>O) for '''2-mL O2k-chamber''':  
:::: '''Preparation of 800 mM stock solution''' (dissolved in H<sub>2</sub>O) for '''2-mL O2k-chamber''':  


::::# Weigh 1.584 g of ascorbate sodium salt and dissolve in 10 mL H<sub>2</sub>O
::::# Weigh 792.4 g of ascorbate sodium salt and dissolve in 3 mL H<sub>2</sub>O
::::# To prevent autoxidation, adjust pH to approx. 6 with ascorbic acid (a 137.6 mg.mL<sup>-1</sup> solution of pH 2).  
::::# To prevent autoxidation, adjust pH to approx. 6 with 1 M HCl (approx. 85 µL).
::::# Divide into 0.2 mL portions.  
::::# Adjust final volume to 5 mL and divide into 0.2 mL portions in dark vials.
::::# Store frozen at -20 °C.  
::::# Store frozen at -20 °C.  


Line 42: Line 41:
:::: '''Preparation of 1000 mM stock solution''' (dissolved in H<sub>2</sub>O) for '''0.5-mL O2k-chamber''':  
:::: '''Preparation of 1000 mM stock solution''' (dissolved in H<sub>2</sub>O) for '''0.5-mL O2k-chamber''':  


::::# Weigh 1.980 g of ascorbate sodium salt and dissolve in 10 mL H<sub>2</sub>O
::::# Weigh 990.5 g of ascorbate sodium salt and dissolve in 3 mL H<sub>2</sub>O
::::# To prevent autoxidation, adjust pH to approx. 6 with ascorbic acid (a 137.6 mg.mL<sup>-1</sup> solution of pH 2).  
::::# To prevent autoxidation, adjust pH to approx. 6 with 1 M HCl.
::::# Divide into 0.2 mL portions.  
::::# Adjust final volume to 5 mL and divide into 0.2 mL portions in dark vials.
::::# Store frozen at -20 °C.  
::::# Store frozen at -20 °C.  



Revision as of 11:15, 22 March 2021


high-resolution terminology - matching measurements at high-resolution


Ascorbate

Description

In respiratory assays for cytochrome c oxidase activity (Complex IV, CIV), ascorbate is added as regenerating system to maintain TMPD in a reduced state. It has to be titrated into the respiration medium prior to the addition of TMPD, otherwise the autoxidation reaction velocity is permanently elevated.

Abbreviation: As

Reference: MiPNet06.06

Application in HRR

As: Ascorbate sodium salt (Sodium L-ascorbate); Sigma A 7631, 100 g, store at RT; M = 198.1 g·mol-1.
Caution: Solutions of ascorbic acid are rapidly oxidized in air and in alkaline media; the compound should be protected from air and light. See 'ascorbid acid' on PubChem


Preparation of 800 mM stock solution (dissolved in H2O) for 2-mL O2k-chamber:
  1. Weigh 792.4 g of ascorbate sodium salt and dissolve in 3 mL H2O
  2. To prevent autoxidation, adjust pH to approx. 6 with 1 M HCl (approx. 85 µL).
  3. Adjust final volume to 5 mL and divide into 0.2 mL portions in dark vials.
  4. Store frozen at -20 °C.
»O2k manual titrations: MiPNet09.12
  • Titration volume (2-mL O2k-chamber): 5 µL using a 25 µL Hamilton syringe.
  • Final concentration: 2 mM.


Preparation of 1000 mM stock solution (dissolved in H2O) for 0.5-mL O2k-chamber:
  1. Weigh 990.5 g of ascorbate sodium salt and dissolve in 3 mL H2O
  2. To prevent autoxidation, adjust pH to approx. 6 with 1 M HCl.
  3. Adjust final volume to 5 mL and divide into 0.2 mL portions in dark vials.
  4. Store frozen at -20 °C.
»O2k manual titrations: MiPNet09.12
  • Titration volume (0.5-mL O2k-chamber): 1 µL using a 10 µL Hamilton syringe.
  • Final concentration: 2 mM.