== Corrigendum ==
 
The polarization voltages for the  H2O2 sensors from WPI is model dependent. For the 2 mm sensor that  usually will be used together with the O2k, the WPI recommendation is  800 to 900 mV, therefore -800 to -900 mV should be set in the O2k, not  -400 mV as stated in MiPNET15.05.
[[User:Fasching Mario|Fasching Mario]] 11:09, 20 January 2014 (CET)