Description | O2k-Barometric Pressure Transducer, A/D converter and digital output to DatLab for continuous recording of barometric pressure [kPa or mmHg], integrated into the air calibration of the OroboPOS (MiPNet06.03 POS-calibration-SOP). Integral component of the O2k-Main Unit. The warranty on the accuracy of the signal obtained from the O2k-Barometric Pressure Transducer expires within three years. |
---|---|
Product ID | 20050 |
Type | O2k-Main Unit |
Link | O2k-Core |
Image |
Accuracy
Quick accuracy check for the barometric pressure transducer 2013-10-21:
- 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa
- nearest metrological station ( Innsbruck airport, alt: 585 m above sea level): pressure = 95.5 kPa
Fasching Mario 14:45, 21 October 2013 (CEST)