O2k-Main Unit

From Bioblast

                



O2k-Main Unit

O2k-Catalogue

Description The O2k-Main Unit is a component of the O2k-Core. The O2k-Main Unit consists of functionally defined, integral elements, the (O2k-Main Basic, O2k-Peltier Temperature Control, two O2k-Electromagnetic Stirrer Control units, two O2k-Amperometric OroboPOS Channels, O2k-Barometric Pressure Transducer), which cannot be obtained separately.
Product ID 20000
Type O2k-Core, Catalogue
Link O2k-Core
Image

The O2k-Main Unit consists of

These are integral elements, not separately available.

_

_

The query description has an empty condition.


O2k-Series

The Oxygraph-2k is continuously evolving. While all O2k instruments have the same high standard for measuring respiration, there are differences with respect to:

the serial number of each O2k is printed on a sticker at the rear of the O2k

Therefore it may be necessary to identify the Series of your oxygraph. The series is specified as the first character of the serial number of the O2k, shown on the label on the rear of the O2k-Main Unit MiPNet12.06. A serial number B-xxxx or C-xxxx denotes O2k-Series B or C, while D-xxxx denotes an O2k-Series D.



on-line the serial number is displayed in the signal line

With DatLab running on-line connected to the O2k, the serial number of the currently connected Oxygraph-2k is displayed:

  • in the right corner of the status line, besides the DatLab version number (bottom).
on-line the serial number is displayed in the signal line
  • in the window caption of the O2k-Control window [F7].

Support of O2k-Modules

The Oxygraph-2k (Series E), or O2k (Series A-D) requires the O2k-MultiSensor electronic upgrade. An exception is the application of an NO sensor in a single chamber of the Series A to C Oxygraph-2k.

Power Supply

The O2k is delivered with one of three different types of power supply cables:

Please note: Use 21111-01 O2k-Main Power Cable\230 V\Europe (Hybrid German/French plug CEE 7-7. Type E+F) only for compatible socket. Note that some danish sockets will also accept the CEE7/7 plugs, but there is no ! grounding connection with these plugs because a male ground pin is required on the plug.

O2k-Main Unit Technical Service

For technical service purposes the term "Main unit" signifies the main part of the oxygraph excluding all removable parts, especially POS (technical service), POS connector (technical service), glass chamber, ... Situated inside the main unit is most of the electronics and the entire heating/cooling and steering system. In case of technical problems this modular approach usually allows to replace/ service just the affected part avoiding to have to send the main unit for repair. To locate any problem to a specific part please follow these instruction:

Locating a problem

Only after OROBOROS Instruments (instruments@oroboros.at) has been contacted for troubleshooting and the problem has been located at the main unit should the main unit be sent to Oroboros instruments for repair. Do not send the main unit without being told to do so by OROBOROS Instruments, since many problems can be solved without sending the main unit. For sending the main unit (or just transporting the oxygraph) follow the instructions for Shipping an oxygraph.