Difference between revisions of "O2k-Open Support alert"

From Bioblast
Line 10: Line 10:
* '''Mar-04 We'''
* '''Mar-04 We'''
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you know that a damaged OroboPOS-Holder might lead to a leaky chamber?'''
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you know that a damaged OroboPOS-Holder might lead to a leaky chamber?'''
::::::: After chamber reassembly you might experience a leaky chamber due to damaged O2k-parts, specially the OroboPOS-Holder - »[[O2k-Chamber_MitoPedia#Leaky_chamber_due_to_damaged_OroboPOS-Holder | Leaky chamber]]«
::::::: After chamber reassembly you might experience a leaky chamber due to damaged O2k-parts, specially the OroboPOS-Holder. - »[[O2k-Chamber_MitoPedia#Leaky_chamber_due_to_damaged_OroboPOS-Holder | Leaky chamber]]«


* '''Mar-11 We'''
* '''Mar-11 We'''

Revision as of 11:41, 4 March 2020


                  


O2k-Open Support

O2k-Open Support alert


Oroboros support system

We provide service and support based on open innovation

- listed in the Oroboros Ecosystem Agenda 2020

Mar 2020

  • Mar-04 We
O2k-Open Support Warning: loss of data Do you know that a damaged OroboPOS-Holder might lead to a leaky chamber?
After chamber reassembly you might experience a leaky chamber due to damaged O2k-parts, specially the OroboPOS-Holder. - » Leaky chamber«
  • Mar-11 We
  • Mar-18 We
O2k-Open Support Info: improve Do you change the slope smoothing to 20 while running experiments?
While running your experiment, especially during uncoupler titrations, it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - »Slope smoothing«


Feb 2020

  • Feb-28 Fr
O2k-Open SupportTake care Calibration parameters are transferred to experimental DatLab files
Save calibration information from selected Marks while DatLab is connected to the O2k during an instrumental calibration run. Air calibration and zero oxygen calibration can be performed and saved in a single DatLab file or in separate DatLab files. - » Oxygen calibration - DatLab«

Jan 2020

  • Jan-31 Fr
O2k-Open SupportTake care Reproducibility committment in high-resolution respirometry: quality control in oxygen calibration
In the DatLab 7.4 software, new instrumental DL-Protocols guide the user automatically step-by-step, to calibrate the instrument accurately. Reproducibility is monitored and documented when following user-friendly standard operating procedures (SOP). - »Oxygen calibration - DatLab«


  • Jan-30 Th
O2k-Open Support Take care Do you check the O2 flux at the end of the air calibration with the O2k-chamber closed?
In DatLab 7.4, instrumental DL-Protocols guide you automatically towards closing the O2k-chamber after air calibration, to monitor "O2 slope neg." (negative) as a test of any biological contamination which would increase oxygen consumption in the absence of a biological sample. - »Oxygen calibration - DatLab«
O2k-Open Support Info: improve Do you set mark ‘R1’ and complete the air calibration correctly?
Your air calibration is not effective until you set mark ‘R1’ and click on Calibrate and copy to clipboard in the O2 calibration window. - »Oxygen calibration - DatLab«


  • Jan-24 Fr
O2k-Open Support Info: improve Do you write the serial number of the polarographic oxygen sensor (POS) in the ‘O2k configuration’ window?
Including the POS serial number in the ´O2k configuration´ window is important to keep a track record of the sensors used according to the O2k. This will be especially helpful if troubleshooting is needed. »O2k configuration«
O2k-Open Support Warning: loss of data Do you perform O2k-Chamber volume calibration?
The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL (or 0.5 mL for the O2k-sV-Module). - »O2k-Chamber MitoPedia«
O2k-Open Support Take care Do you use the correct O2 solubility factor (FM) in the ‘O2 calibration’ window in DatLab?
The O2 solubility factor varies according to the medium used and influences the air calibration. - »Oxygen solubility factor«