Currently Answered Questions from the »Oroboros Ecosystem Agenda«



Apr 2020

  • Apr-14 Tu
O2k-Open SupportTake care Why are the oxygen levels different between chambers prior to my experimental run?
After switching on the O2k, oxygen concentration values might be atypical or different between chambers if air calibration is not performed - » Oxygen calibration - troubleshooting«
  • Apr-07 Tu
O2k-Open Support Info: improve Did you know that you can edit the background parameters in DatLab 7.4 O2 analysis Excel template?
If you forgot to apply the background correction before running your experiment, you can add the values directly to the O2 analysis DL7.4 Excel template and complete your data analysis following the quality control step 2. »Oxygen flux analysis - DatLab 7.4«
  • Apr-02 Th
MitoPedia: SUITDo you want to search the best SUIT protocol for your research questions? Use the SUITbrowser: »SUITbrowser«

Mar 2020

  • Mar-30 Mo
O2k-Open Support Info: improve DatLab 7.4 offers the flexibility to edit a DL-Protocol: »Enable DL-Protocol editing«


  • Mar-25 We
O2k-Open Support Info: improveO2 flux analysis with DatLab 7.4. Use our latest O2 analysis template: »O2k-Videosupport: O2 flux analysis with DatLab 7.4«


  • Mar-23 Mo
New O2k-Open Support Take care Do you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?
(1) By adjusting the dithionite concentration to 2.5 mM, the instrumental O2 background can be performed with the "BG Feedback" settings. - »O2k-sV-Module«
(2) By adjusting the TIP2k setup with specific settings for the O2k-sV-Module, the instrumental O2 background can be performed in the O2k-sV-Module. - »O2k-sV-Module«


  • Mar-11 We
New O2k-Open Support Info: improve Do you change the slope smoothing to 20 while running experiments with biological sample?
While running your experiment, especially during uncoupler titrations, it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - »Slope smoothing«
  • Mar-04 We
OroboPOS-Holder002.jpg


New O2k-Open Support Warning: loss of data A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.
See how you can check that the two blue OroboPOS-Holders are intact. - »O2k-OroboPOS-Holder«



Feb 2020

  • Feb-28 Fr
New O2k-Open SupportTake care Calibration parameters are transferred to experimental DatLab files
Save calibration information from selected Marks while DatLab is connected to the O2k during an instrumental calibration run. Air calibration and zero oxygen calibration can be performed and saved in a single DatLab file or in separate DatLab files. - » Oxygen calibration - DatLab«

Jan 2020

  • Jan-31 Fr
O2k-Open SupportTake care Reproducibility committment in high-resolution respirometry: quality control in oxygen calibration
In the DatLab 7.4 software, new instrumental DL-Protocols guide the user automatically step-by-step, to calibrate the instrument accurately. Reproducibility is monitored and documented when following user-friendly standard operating procedures (SOP). - »Oxygen calibration - DatLab«


  • Jan-30 Th
O2k-Open Support Take care Do you check the O2 flux at the end of the air calibration with the O2k-chamber closed?
In DatLab 7.4, instrumental DL-Protocols guide you automatically towards closing the O2k-chamber after air calibration, to monitor "O2 slope neg." (negative) as a test of any biological contamination which would increase oxygen consumption in the absence of a biological sample. - »Oxygen calibration - DatLab«
O2k-Open Support Info: improve Do you set mark ‘R1’ and complete the air calibration correctly?
Your air calibration is not effective until you set mark ‘R1’ and click on Calibrate and copy to clipboard in the O2 calibration window. - »Oxygen calibration - DatLab«


  • Jan-24 Fr
O2k-Open Support Info: improve Do you write the serial number of the polarographic oxygen sensor (POS) in the ‘O2k configuration’ window?
Including the POS serial number in the ´O2k configuration´ window is important to keep a track record of the sensors used according to the O2k. This will be especially helpful if troubleshooting is needed. »O2k configuration«
O2k-Open Support Warning: loss of data Do you perform O2k-Chamber volume calibration?
The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL (or 0.5 mL for the O2k-sV-Module). - »O2k-Chamber MitoPedia«
O2k-Open Support Take care Do you use the correct O2 solubility factor (FM) in the ‘O2 calibration’ window in DatLab?
The O2 solubility factor varies according to the medium used and influences the air calibration. - »Oxygen solubility factor«