Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "OroboPOS-Connector"

From Bioblast
Line 4: Line 4:
Two units of this item are standard components of the [[O2k-Assembly Kit]] ([[O2k-FluoRespirometer]]), mounted on the [[OroboPOS-Holder]]s at the [[O2k-Main Unit]].
Two units of this item are standard components of the [[O2k-Assembly Kit]] ([[O2k-FluoRespirometer]]), mounted on the [[OroboPOS-Holder]]s at the [[O2k-Main Unit]].
|product id=22200-01
|product id=22200-01
|product type=O2k, O2k-Assembly Kit, OroboPOS, O2k-Respirometer, Startup-O2k
|product type=O2k, O2k-Assembly Kit, OroboPOS, O2k-FluoRespirometer, Startup-O2k
|info=[[O2k-Assembly Kit]], [[MiPNet19.18A O2k-start|O2k-Manual]]
|info=[[O2k-Assembly Kit]], [[MiPNet19.18A O2k-start|O2k-Manual]]
|product image=[[Image:POS connector and cable connection.jpg|right|180px]]
|product image=[[Image:POS connector and cable connection.jpg|right|180px]]

Revision as of 14:34, 10 August 2017

                



OroboPOS-Connector

O2k-Catalogue

Description OroboPOS-Connector (blue POM), with male connection to OroboPOS sensor head (POS) and with cable and male plug fitting into O2k-Main Unit.

Two units of this item are standard components of the O2k-Assembly Kit (O2k-FluoRespirometer), mounted on the OroboPOS-Holders at the O2k-Main Unit.

Product ID 22200-01
Type O2k, O2k-Assembly Kit, OroboPOS, O2k-FluoRespirometer, Startup-O2k
Link O2k-Assembly Kit, O2k-Manual
Image
POS connector and cable connection.jpg


MitoPedia O2k and high-resolution respirometry: O2k hardware 



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Functions

The OroboPOS-Connector has several different functions:
  • It mechanically presses the POS against the glass chamber, ensuring a tight fit.
  • It connects the POS electronically to the Main unit transmitting the signal.
  • In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the Main unit. For this reason OroboPOS connectors for Series A-C and for Series D are NOT interchangeable.
The OroboPOS-Connector is sensitive to damage by ESD when disconnected from the O2k-Main Unit and when the POS head is removed. Further details: MiPNet14.01_ESD-damage.

Cleaning

When experimental media or electrolyte (during service of the polarographic oxygen sensor OroboPOS leak into the mechanical part of the OroboPOS-Connector, operation of the spring mechanism may be restricted. In such cases, the body of the OroboPOS-Connector should be washed with distilled water. To do so, screw the OroboPOS head onto the OroboPOS-Connector. Make sure that the plug connecting to the O2k-Main Unit does not get wet (protect with parafilm). Rinse the body of the OroboPOS-Connector with distilled water, pulling the spring-loaded part at the base several times. Salt crystals and media clogging the mechanical part are thus washed out. Wipe off the water while again pulling down the spring-loaded part.

Cleaning the electrical connections

Unscrew the OroboPOS and check the electrical connections (the gold pin and thread on the OroboPOS and OroboPOS-Connector). Remove any contaminating grease and moisture with a fine paper cloth. Soak a paper cloth in ethanol and carefully clean the gold pin and threads. If the gold pin and thread on the POS connector are discoloured or appear stained, clean repeatedly with a paper towel soaked in water and subsequently in ethanol, until no more traces appear on the paper towel used for cleaning. If the O2k is used in a humid environment, apply a drop of contact oil from the OroboPOS-Service Kit to each of the golden pins and threads on POS connector and POS. Perform an Oxygen sensor test as a quality control of proper OroboPOS-connector and sensor performance.

Replacing the OroboPOS-Connector

If you need to replace the Orobo-POS Connector, an OroboPOS-Connector Service may be possible.