(40 intermediate revisions by 12 users not shown) |
Line 1: |
Line 1: |
| {{Product | | {{Product |
| |description='''OroboPOS-Connector''' (blue POM), with male connection to [[OroboPOS]] sensor head (POS) and with cable and male plug fitting into [[O2k-Main Unit]]. | | |description='''OroboPOS-Connector''' (blue POM), with male connection to [[OroboPOS]] sensor head (POS) and with cable and male plug fitting into [[O2k-Main Unit]].Two units of this item are standard components of the [[O2k-Assembly Kit]] ([[O2k-FluoRespirometer]]), mounted on the [[OroboPOS-Holder]]s at the [[O2k-Main Unit]]. |
| Β | |
| [[O2k-Core]]: 2 in [[O2k-Assembly Kit]], mounted on the [[OroboPOS-Holder]]s at the [[O2k-Main Unit]]. | |
| |product id=22200-01 | | |product id=22200-01 |
| |product type=O2k, O2k-Assembly Kit, OroboPOS | | |product type=O2k, OroboPOS, O2k-FluoRespirometer, Power O2k-FluoRespirometer, Power O2k-Respirometer, Startup O2k-Respirometer |
| |info=[[O2k-Assembly Kit]], [http://www.oroboros.at/?O2k-Start O2k-Manual @OROBOROS] | | |info=[[O2k-Assembly Kit]], [[MiPNet22.11 O2k-FluoRespirometer manual|O2k-Manual]], [[OroboPOS-Connector MitoPedia]] |
| |product image=[[Image:POS connector and cable connection.jpg|right|180px]] | | |product image=[[Image:POS connector and cable connection.jpg|right|180px]] |
| }} | | }} |
|
| |
|
| |
| The two electronic connections of the POS connector (to the [[POS]] and to the [[Main unit]]) make a POS connector that has been disconnected from the [[POS]] or from the [[Main unit]] or from both particular sensitive to damage by [[ESD]]. '''Therefore it is of special importance to follow the guidelines for ESD protection [[MiPNet14.01_ESD_Damage]] whenever handling a POS connector.'''
| |
|
| |
| == Functions ==
| |
| The POS connector has several different functions:
| |
| * It mechanically presses the [[POS]] against the [[glass chamber]], ensuring a tight fit.
| |
| * It connects the [[POS]] electronically to the [[main unit]] transmitting the signal.
| |
| * In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the [[Main unit]]. For this reason POS connectors for Series A-C and for Series D are NOT interchangeable.
| |
|
| |
| == Technical Service ==
| |
| Nearly all problems usually associated with the [[POS]] may in fact be caused by the POS connectors, e.g.
| |
| *unstable O2 signal
| |
| *unstable O2 flux
| |
| *high O2 signal at air saturation
| |
| *high O2 signal at zero oxygen
| |
|
| |
|
| |
| === Troubleshooting procedure ===
| |
|
| |
| Confirm the problem by doing a [[Sensor test]] in the absence of biological material
| |
|
| |
| If no no problem was visible in the [[Sensor test]] check the following [[DatLab]] settings:
| |
| * [[Gain (O2 channel)]]
| |
| * [[POS calibration]]
| |
| * Scaling
| |
|
| |
| If the problem is not visible when observing the [[Raw signal]] but is visible when observing the calibrated oxygen signal, then there is probable a problem with the [[POS calibration]].
| |
|
| |
| If the problem was visible in the [[Sensor test]]:
| |
| * Proceed to [[ Locating a problem|Locate the Problem ]]
| |
| * If thereby the problem was confirmed to be caused the OroboPOS-Connector: Follow the instructions for cleaning and applying [[contact oil]] as described in the sensor service manual [http://www.oroboros.at/index.php?id=o2k-pos-service [MiPNet08.04]] and in more detail at [[Cleaning the POS connector]]. If you have also done a [[Sensor service]] or changed the [[POS membranes]] allow the oxygraph to run over night before repeating the [[Sensor test]]. If all service precautions fail to solve the problem, or a new POS connector has to be applied.
| |
| * If the problem was localized on an other component follow the instructions for this component. If you could not locate the problem contact OROBOROS Instruments.
| |
|
| |
| === OroboPOS-Connector Technical Service Pages ===
| |
| {{#ask: mainlabel=Title|[[Category:All]] | [[Technical service::POS connector]]| ?Technical service}}
| |
|
| |
| {{#set: Technical service=POS connector|Technical service=O2 signal}}
| |
|
| |
| [[Category:Technical service]]
| |