Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

OroboPOS-Connector

From Bioblast

                



OroboPOS-Connector

O2k-Catalogue

Description OroboPOS-Connector (blue POM), with male connection to OroboPOS sensor head (POS) and with cable and male plug fitting into O2k-Main Unit.

O2k-Core: 2 in O2k-Assembly Kit, mounted on the OroboPOS-Holders at the O2k-Main Unit.

Product ID 22200-01
Type O2k, O2k-Assembly Kit, OroboPOS
Link O2k-Assembly Kit, O2k-Manual @OROBOROS
Image
POS connector and cable connection.jpg


The two electronic connections of the POS connector (to the POS and to the Main unit) make a POS connector that has been disconnected from the POS or from the Main unit or from both particular sensitive to damage by ESD. Therefore it is of special importance to follow the guidelines for ESD protection MiPNet14.01_ESD_Damage whenever handling a POS connector.

Functions

The POS connector has several different functions:

  • It mechanically presses the POS against the glass chamber, ensuring a tight fit.
  • It connects the POS electronically to the main unit transmitting the signal.
  • In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the Main unit. For this reason POS connectors for Series A-C and for Series D are NOT interchangeable.

Technical Service

Nearly all problems usually associated with the POS may in fact be caused by the POS connectors, e.g.

  • unstable O2 signal
  • unstable O2 flux
  • high O2 signal at air saturation
  • high O2 signal at zero oxygen


Troubleshooting procedure

Confirm the problem by doing a Sensor test in the absence of biological material

If no no problem was visible in the Sensor test check the following DatLab settings:

If the problem is not visible when observing the Raw signal but is visible when observing the calibrated oxygen signal, then there is probable a problem with the POS calibration.

If the problem was visible in the Sensor test:

  • Proceed to Locate the Problem
  • If thereby the problem was confirmed to be caused the OroboPOS-Connector: Follow the instructions for cleaning and applying contact oil as described in the sensor service manual [MiPNet08.04] and in more detail at Cleaning the POS connector. If you have also done a Sensor service or changed the POS membranes allow the oxygraph to run over night before repeating the Sensor test. If all service precautions fail to solve the problem, or a new POS connector has to be applied.
  • If the problem was localized on an other component follow the instructions for this component. If you could not locate the problem contact OROBOROS Instruments.

OroboPOS-Connector Technical Service Pages

The query description has an empty condition.


... more about "OroboPOS-Connector"
OroboPOS-Connector (blue POM), with … OroboPOS-Connector (blue POM), with male connection to OroboPOS sensor head (POS) and with cable and male plug fitting into O2k-Main Unit.</br></br>O2k-Core: 2 in O2k-Assembly Kit, mounted on the OroboPOS-Holders at the O2k-Main Unit.O2k-Main Unit. +
22200-01 +
O2k +, O2k-Assembly Kit +  and OroboPOS +
POS connector +  and O2 signal +