|description=The '''POS holder''' is screwed into the copper block of the O2k, fixing the glass chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.
__TOC__
|product id=25100-21
{{MitoPedia without banner
|product type=O2k
|abbr=
|info=MiPNet12.06
|description=[[Image:Pos_holder.JPG|right|180px]]The '''OroboPOS-Holder''' (POS-Holder), made from blue POM, is screwed into the copper block of the [[O2k-Main Unit]], guiding the [[OroboPOS-Connector]] with the [[OroboPOS]] sensor head (POS) to the 2-mL [[O2k-chamber]], and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-chamber with the [[OroboPOS-Seal Tip]]. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.
 
Two units of this item are standard components mounted on the [[O2k-Main Unit]].
|info=[https://www.oroboros.at/index.php/product/orobopos-holder/ '''Product details and purchase information''']
 
}}
== Assembly ==
 
::::* Manual: [https://wiki.oroboros.at/images/2/29/MiPNet22.11_O2k-FluoRespirometer-manual.pdf MiPNet22.11_O2k-FluoRespirometer-manual.pdf] Section 5.2.
 
::::* Video: [https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]
== A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber ==
 
:::* The OroboPOS-Holder may be damaged:
 
::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (''see center picture below''). 
::::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]].
 
::::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see'' pictures below).
 
[[File:OroboPOS-Holder002.jpg|left|200px]]
[[File:OroboPOS-Holder003.jpg|right|200px]]
[[File:OroboPOS-Holder004.jpg|centre|200px]]
 
* Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
* Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged.
* Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.
 
 
::::* Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: [https://www.oroboros.at/index.php/o2k-products/spares/ Spare parts & consumables]
''See'' [[Talk:OroboPOS-Holder|Discussion page]]
 
== Keywords ==
{{Template:Keywords: Chamber volume}}
 
 
 
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry
}}
}}

Revision as of 18:05, 10 November 2020


                  


O2k-Open Support

POS-Holder



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

Pos holder.JPG

The OroboPOS-Holder (POS-Holder), made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding the OroboPOS-Connector with the OroboPOS sensor head (POS) to the 2-mL O2k-chamber, and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-chamber with the OroboPOS-Seal Tip. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.

Two units of this item are standard components mounted on the O2k-Main Unit.


Reference: Product details and purchase information

Assembly

A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber

  • The OroboPOS-Holder may be damaged:
  1. During chamber assembly, when fixing the chamber in a wrong rotational position with the O2k-Chamber Holder (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (see center picture below).
  2. After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the O2k-Main Unit.
  • We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (see pictures below).
OroboPOS-Holder002.jpg
OroboPOS-Holder003.jpg
OroboPOS-Holder004.jpg
  • Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
  • Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged.
  • Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.


See Discussion page