Difference between revisions of "Oxygen calibration - DatLab"

From Bioblast
(35 intermediate revisions by 7 users not shown)
Line 1: Line 1:
{{MitoPedia
{{Technical support}}
|description='''O2 calibration''' is the calibration in DatLab of the oxygen sensor. It is a prerequisite for obtaining accurate measurements of respiration. Accurate calibration of the oxygen sensor depends on (1) equilibration of the incubation medium with air oxygen partial pressure at the temperature defined by the experimenter; (2) zero oxygen calibration; (3) high stability of the POS signal tested for sufficiently long periods of time; (4) linearity of signal output with oxygen pressure in the range between oxygen saturation and zero oxygen pressure; and (5) accurate oxygen solubility for aqueous solutions for the conversion of partial oxygen pressure into oxygen concentration. The standard oxygen calibration procedure is described below for high-resolution respirometry with the automatic calibration routine by DatLab.^
{{MitoPedia without banner
|description='''O2 calibration''' is the calibration in DatLab of the oxygen sensor. It is a prerequisite for obtaining accurate measurements of respiration. Accurate calibration of the oxygen sensor depends on (''1'') equilibration of the incubation medium with air oxygen partial pressure at the temperature defined by the experimenter; (''2'') zero oxygen calibration; (''3'') high stability of the POS signal tested for sufficiently long periods of time; (''4'') linearity of signal output with oxygen pressure in the range between oxygen saturation and zero oxygen pressure; and (''5'') accurate oxygen solubility for aqueous solutions for the conversion of partial oxygen pressure into oxygen concentration. The standard oxygen calibration procedure is described below for high-resolution respirometry with the calibration routine using instrumental calibration DL-Protocols in [[DatLab]].
|info=[[MiPNet06.03 POS-calibration-SOP]], [[MiPNet12.20_O2k-calibration_tutorial]]
|info=[[MiPNet06.03 POS-calibration-SOP]], [[MiPNet12.20_O2k-calibration_tutorial]]
}}
}}
[[File:Calibration-menu.png|right|200px]]
[[File:Calibration-menu.png|right|300px]]
{{MitoPedia methods
|mitopedia method=Respirometry
}}
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=DatLab
}}
__TOC__
__TOC__


== Air calibration - general description ==
== Oxygen calibration in instrumental and experimental DatLab files ==
 
[[File:Calibration window steps.png|right|500px]]
:::: Air saturation is achieved by stirring the aqueous medium in contact with air in the O2k-chamber without sample, following the procedures below.
::: '''Instrumental calibration'''
:::: It is common practice to save instrumental calibration tests in separate DatLab files, which include (''1'') air calibration and zero oxygen calibration in the same calibration file, or (''2'') air calibration and zero oxygen calibration in different calibration files. '''Step <span style="background:red">1</span>''': Select calibration sections from [[Marks - DatLab |Marks]] (R1, R0) while DatLab is connected to the O2k during an instrumental calibration run. '''Step <span style="background:red">2</span>''': Click on <span style="background:lightgrey">Calibrate and copy to clipboard</span> in the O2 calibration window (in menu 'Calibration' > 'Oxygen, O2'). '''Step <span style="background:red">3</span>''': Click on <span style="background:lightgrey">Save and disconnect</span> in menu File. The calibration file is saved, and DatLab disconnects from the O2k.  


[[File:Stopper-spacer.png|thumb|right|200px|'''Stopper-spacers''' used for air calibration with white PVDF stopper (Chamber A) or black PEEK stopper (Chamber B).]]
::: '''Calibration parameters are automatically transferred to experimental DatLab files'''
:::: '''Step <span style="background:red">4</span>''': When a new DatLab file is opened any time later, the latest calibration parameters are automatically applied in the new DatLab file — independent of maintaining the O2k in operation or switching it off, and independent of maintaining DatLab active or re-opening it after exit.


1. Add incubation medium into the chambers with an excess volume of at least 0.1 ml above the experimental chamber volume (2 ml) in order to fill the O2k-chamber and injection capillary of the stopper when it is fully inserted (closed). The volume does not have to be accurate, as long as it is above the minimum volume. Switch on the stirrers either during or after addition of the medium.
::: '''A calibration can be performed and calibration parameters saved during an experimental run'''
:::: In the specific case of an experiment that involves an aerobic-anoxic transition, zero oxygen calibration is performed within the experimental file, setting a Mark (R0) on the oxygen signal in the anoxic section, and clicking on <span style="background:lightgrey">Calibrate and copy to clipboard</span> in the O2 calibration window (in menu 'Calibration' > 'Oxygen, O2').


2. Insert the stoppers slowly to their volume-calibrated position. Suck off excess medium ejected through the injection capillary and collected in the well of the stopper. Then lift the stoppers using the stopper-spacer tool, leaving a gas volume above the liquid phase for final air equilibration - [[Open chamber]] position.
::: '''Calibration parameters can be selectively imported from any DatLab file'''
:::: '''Step <span style="background:red">5</span>''': In addition, the calibration parameter for air calibration, zero oxygen calibration, or both can be imported from a selected DatLab file. This is particularly important, if a zero calibration is performed after completion and saving an experimental file. '''Step <span style="background:red">6</span>''': A single calibration value (R1 or R0) can be edited manually. Take '''Step <span style="background:red">7</span>''' to apply the selected calibration parameters.


The central level of the gas phase remains above the rotating stirrer bar, preventing bubbles and foam from being formed which would block gas exchange. To ensure a well defined pO2 in the gas phase, the gas volume has to be renewed (exchanged for air), if the medium was originally not near air saturation. This is achieved simply by fully inserting and re-opening the stopper. Equilibration is a slow process: stability should be reached within one hour (figure to the right). A stirrer test F9 can be performed during equilibration.


3. After stabilization of the POS signal, the recorded signal at air saturation, R1, is about 1-3 V at Gain 1 and a temperature of 25-37°C. A signal of 1 V corresponds to a signal current of the POS of 1 µA (corresponding to 2 V at Gain 2). Under all experimental conditions, the raw signal must be <10 V.
== Instrumental DL-Protocols ==


Continue recording for 3-10 min to check for signal stability. You may proceed at this point with an O2 background test (see below).
[[Image:Logo OROBOROS-DatLab.jpg|left|60px||link=http://wiki.oroboros.at/index.php/DatLab |DatLab]] Instrumental DL-Protocols guide the O2k operator step-by-step through the calibration procedure, adding automatically the event names and mark names set during the calibration run.


[[File:Air calibration.png|thumb|right|600px|'''Air calibration:''' Oxygen concentration (blue plot; 180 µM; full scale 50 nmol/ml or 50 µM) over 1 h after switching on the Oxygraph-2k (Power-O2k P6, chamber A) and setting the experimental temperature at 37 °C, using medium stirred for equilibration with a gas phase of air at 575 m altitude. The red plot is the negative slope of oxygen concentration over time [pmol/(s­*ml)] on the right Y-axis, with zero in the middle position. A slope of zero (for uncorrected ‘O2 Slope neg.’) indicates a constant O2 signal over time. (2014-07-24 P4-01.DLD)]]
[[File:DLP O2-calibration air.png|right|800px]]
[[File:DLP O2-calibration air-J01.png|right|800px]]
=== Air calibration: O2-calibration_air.DLP ===


== Air calibration in DatLab ==
::::* O2 sensor test and calibration at air saturation: Performed routinely every day before starting measurements.


'''1. Select Graph layout:'''
::: Event <span style="background:yellow">E: Stirrer test*</span style>
:::: 1 - Perform a stirrer test [F9], even before final equilibration.
:::::* Both stirrers are switched off automatically for 30 s (default).
:::::* Re-start of the stirrers is automatic (automatic event: On).
:::::* The increase of the O2 signal should be rapid and monoexponential.
:::: 2 - Proceed with mark R1.


Go to menu "Layout", check "O2" and from "Standard layouts" select "01 Calibration show Temp".
::: Mark M: R1 - air calibration
This is typically the first layout used after switching on the O2k. Oxygen concentration (blue lines, left Y-axis) and O2 slope (not corrected for instrumental background; red lines, right Y-axis) are displayed on the top graph for the left chamber, and below for the right chamber. The third graph (bottom) shows the block temperature on the left Y axis and the Peltier power on the right Y axis. Only when both temperature and Peltier power are constant, the chambers have reached thermal equilibrium. The next step is to observe equilibration of the oxygen signal with a defined gas phase above the stirred aqueous phase ('open' chamber; usually with air as the first step) to perform an oxygen calibration.
:::: 1 - After the stirrer test, the signal should stabilize within 40 min.
:::::* O2 slope neg. should approach zero, at ± 1 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup>.
:::::* Noise: An 'O2 slope neg.' within ± 2 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup> is optimal (± 4 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup> is acceptable)
:::: 2 - Set the marks R1 for each chamber on 'Y1: O2 concentration'.


'''2. Mark'''
:::: 3 - Open the O2 calibration window in menu 'Calibration' > 'Oxygen, O2'.
:::: 4 - Select Mark R1 for 'Air calibration'.
:::::* 'POS signal: Recorded' should be 1 to 3 V at 25 to 37 °C (up to 1000 m above sea level; pb 101 to 90 kPa) at gain = 1.
:::::* 'O2 slope neg.' should be within the range of -1 to 1 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup>.
:::: 5 - Calibrate and copy to clipboard.
:::: 6 - Paste into the Excel file: "Template O2 calibration.xlsx".
:::::* "Template O2 calibration.xlsx" is provided on the USB flash-drive (folder 'Oroboros O2k-Course on HRR\O2k\DL-Protocols') or can be opened from DatLab: [[Run_DL-Protocol/Set_O2_limit#Instrumental:_Browse_DL-Protocols_and_templates]]
:::: 7 - Proceed with event 1C.


Mark a section of the experiment at air saturation, when signal stability is reached. This should be done real-time to save default calibration information. Corrections are possible after disconnecting from the O2k. For calibration, follow steps (1) to (5):
::: Event <span style="background:yellow">E: 1C</span style> - Close chamber
::::: 1 - After air calibration, close the chamber by gently pushing the stopper fully downwards.
::::: 2 - Siphon off excess medium from the top of the stopper using the Integrated Suction System (ISS).
::::: 3 - Check that any air bubbles are removed from the chamber.
::::: 4 - Switch to 'Y2: O2 slope neg.' by clicking on the Y2 label on the graph legend.
::::: 5 - Proceed with mark J°1.


1. Select a graph by a click (left mouse button) into the graph or directly by step 2.
::: Mark M: J°1 - O2 BG flux ~ air saturation (instrumental background flux after air calibration)
::::: 1 - Set a mark (J°1) after event 1C at a section, where the O2 slope is stable, and observe the signal.
:::::* 'O2 slope neg.' should be within the range of ± 2 to 4 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup>.
:::::* 'O2 slope neg.' > 4 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup> may indicate a biological contamination in the medium or O2k-chamber.
::::: 2 - Save the file in menu 'File' > 'Save and disconnect' before starting an experiment.
:::::* Although the calibration file is closed, the values will be automatically used as default calibration values for the next measurements.


2. Select the oxygen signal as the active plot by a click on Y1 in the figure legend on the right of the graph. The active plot is highlighted.
[[File:DLP O2-calibration air and zero.png|right|800px]]
[[Image:Logo OROBOROS-DatLab.jpg|left|60px||link=http://wiki.oroboros.at/index.php/DatLab |DatLab]]
=== Air and zero calibration: O2-calibration_air and zero.DLP ===


3. Only if ‘Mouse Control: Zoom’ mode has been activated: Select “Mouse Control: Mark” in the Graph menu or press Ctrl+M.
::::* O2 sensor test and calibration at air saturation and zero oxygen:
::::: This DL-Protocol is identical in the section of air calibration, and continues with zero calibration as follows:


4. Set a mark: Hold Shift and click the left mouse button, move the cursor along the time axis, release the left mouse button at the end of the section to be marked.  
::: Mark M: J°1 - O2 BG flux ~ air saturation (instrumental background flux after air calibration)
Remove a section of the mark or the total mark: Move the cursor with Shift+right-click along the time axis, release the mouse button at the end of the section of the mark to be deleted.
::::: 1 - Set a mark (J°1) after event 1C at a section, where the O2 slope is stable, and observe the signal.
:::::* 'O2 slope neg.' should be within the range of 2 to 4 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup>.
:::::* 'O2 slope neg.' > 4 pmol∙s<sup>−1</sup>∙mL<sup>−1</sup> may indicate a biological contamination in the medium or O2k-chamber.
::::: 2 - Switch to plot 'Y1: O2 concentration' by clicking on the Y1 label on the graph legend.
::::: 3 - Proceed with event 2Dith*.


5. Rename the mark: Left mouse click on the bar of the mark. Rename the mark for air calibration as “R1” (and the mark for zero calibration as “R0”).
::: Event <span style="background:yellow">E: 2Dith*</span style> - Manual dithionite titration
::::: 1 - Events 2Dith*: Inject fresh Na-dithionite solution ("zero solution") until the O2 signal is reduced close to 0 µM.
:::::* A few minutes are required until a stable "zero" signal is obtained     
:::::* Repeat until the signal does not respond by a further decline. Siphon off excess medium from the stopper.
::::: 2 - Switch to plot 'Y1: O2 concentration' by clicking on the Y1 label on the graph legend.
::::: 3 - Proceed with mark R0.  


'''3. Calibrate'''
::: Mark M: R0 - Zero O2 calibration
[[File:Calibration window.PNG|thumb|right|600px|'''Calibration window''']]
::::: 1 - Set a mark (R0) after the last event 2Dith* when the signal is stable (Y1: O2 concentration).
::::: 2 - Open the O2 calibration window in menu 'Calibration' > 'Oxygen, O2'.
::::: 3 - Select mark R0 for 'Zero calibration'.
:::::* R0 should be stable and < 2% of the signal at air saturation, but < 5% is acceptable.
::::: 4 - Select mark R1 for 'Air calibration'.
::::: 5 - Calibrate and copy to clipboard.
::::: 6 - Paste into the Excel file: "Template O2 calibration.xlsx".
:::::* "Template O2 calibration.xlsx" is provided on the USB flash-drive (folder 'Oroboros O2k-Course on HRR\O2k\DL-Protocols') or can be opened from DatLab: [[Run_DL-Protocol/Set_O2_limit#Instrumental:_Browse_DL-Protocols_and_templates]]
::::: 7 - Save the file in menu 'File' > 'Save and disconnect' before starting the next file.
:::::* After the calibration file is saved while connected to the O2k, the calibration parameters will be automatically used as default after restarting DatLab.


1. Open the DatLab calibration window by double clicking on O2 calib. for O2k-chamber A or B in the O2k-status line (bottom left and right). Alternatively, press F5 to open the calibration window for the active plot.


2. Left click on the pull down button and select the appropriate mark (R1). The average voltage (Raw signal [V]) recorded over the marked section is shown in the corresponding field on the right. The corresponding signal stability is displayed as the uncorrected negative slope of the signal during calibration in [pmol/(s­.ml)]. Temperature and barometric pressure are displayed as measured over the marked section.
== Oxygen calibration - general description ==
Calibration values R1 and R0 can be edited numerically, without exerting an influence on c1. If temperature or barometric pressure are edited, then c1 is recalculated for the changed conditions.


3. If not set previously, enter the oxygen solubility factor of the medium, FM, relative to pure water. For more info on this factor: MiPNet06.03 POS-calibration-SOP. For documentation purposes, enter the name of the experimental incubation medium in the coresponding field.
:::: The general description lists the calibration steps in further detail, without considering the convenience of using the instrumental DL-Protocols listed above.


4. Left click on "Calibrate and copy to clipboard". After clicking on this button all changes in the entire calibration window are applied. The entire plot of oxygen concentration is re-calibrated [µM = nmol/ml], and the corresponding negative slope or volume-specific oxygen flux [pmol/(s.ml)] is now based on this new calibration. Calibration parameters are automatically copied to clipboard for entry into the spreadsheet O2 calibration.xlsx.
:::: DatLab converts the [[oxygen signal]] of the polarographic oxygen sensor (OroborPOS or POS) to oxygen concentration during calibration comprising an [[air calibration]] (R1) and a [[zero calibration]] (R0).
:::# Appropriate calibration conditions are obtained in the O2k-chamber at experimental temperature.
:::# R1 and R0 are marked on the (blue) plot of oxygen concentration.
:::# The marks for R1 and R0 are pulled down in the <span style="background:lightgrey">O2 calibration</span> window of DatLab.
:::# Click on <span style="background:lightgrey">Calibrate and copy to clipboard</span> before closing the calibration window.


:::: Frequently, assumed hardware problems turn out to be a simple calibration problem.


== Zero calibration - general description ==
:::: Air saturation is achieved by stirring the aqueous medium in contact with air in the O2k-chamber without sample.


Zero oxygen calibration may be achieved by different means. Routinely, following chamber assembly or exchange of the POS membrane, zero solution (Na-dithionite, OroboPOS-Service Kit) is titrated into the chamber, which is part of the automatic TIP2k-supported instrumental background test (MiPNet14.06).  
[[File:Stopper-spacer.png|thumb|right|200px|'''Stopper-spacers''' used for air calibration with white PVDF stopper (Chamber A) or black PEEK stopper (Chamber B).]]


Alternatively, zero calibration can be done with mitochondria or cell suspensions that allow complete oxygen depletion.
:::# Add incubation medium into the chambers with an excess volume of at least 0.1 mL above the experimental chamber volume (2 mL or 0.5 mL) in order to fill the O2k-chamber and injection capillary of the stopper when it is fully inserted (closed). The volume does not have to be accurate, as long as it is above the minimum volume. Switch on the stirrers either during or after the addition of the medium.
:::# Insert the stoppers slowly to their volume-calibrated position. Suck off excess medium ejected through the injection capillary and remaining in the well of the stopper. Then lift the stoppers using the stopper-spacer tool, leaving a gas volume above the liquid phase for final air equilibration - [[Open chamber]] position. The central level of the gas phase remains above the rotating stirrer bar, preventing bubbles and foam from being formed which would block gas exchange. To ensure a well-defined ''p''<sub>O2</sub> in the gas phase, the gas volume has to be renewed (exchanged for air), if the medium was originally not near air saturation. This is achieved simply by fully inserting and re-opening the stopper. Equilibration is a slow process: stability should be reached within one hour (figure to the right). A stirrer test [F9] is performed early during equilibration.
:::# After stabilization of the POS signal, the recorded signal at air saturation, R1, is about 1-3 V (at Gain 1) and a temperature of 25-37 °C. A signal of 1 V corresponds to a signal current of the POS of 1 µA (corresponding to 2 V at Gain 2). Under all experimental conditions, the raw signal must be <10 V. Continue recording for 3-10 min to check for signal stability. You may proceed at this point with an O<sub>2</sub> background test (see below).


== Zero calibration in DatLab ==
[[File:Air calibration.png|thumb|right|600px|'''Air calibration:''' Oxygen concentration (blue plot; 180 µM; scale from 0 to 250 nmol/mL or 0 to 250 µM) over 1 h after switching on the O2k (Power-O2k P4, chamber A) and setting the experimental temperature at 37 °C, using medium stirred for equilibration with a gas phase of air at 575 m altitude. The red plot is the negative slope of oxygen concentration over time [pmol∙s<sup>−1</sup>∙mL<sup>−1</sup>] on the right Y-axis, with zero in the middle position. A slope of zero (for uncorrected ‘O2 Slope neg.’) indicates a constant O<sub>2</sub> signal over time. (2014-07-24 P4-01.DLD)]]


'''1. Mark'''
:::: Zero oxygen calibration may be achieved by chemical or biological depletion of oxygen. (''1'') Chemical: Following chamber assembly or exchange of the POS membrane, zero solution (Na-dithionite, OroboPOS-Service Kit) is titrated into the chamber, which is part of the automatic TIP2k-supported instrumental background test (MiPNet14.06). (''2'') Biological: Zero calibration can be performed with mitochondria or cell suspensions that allow complete oxygen depletion.


Mark a section of the experiment after full depletion of oxygen, when signal stability is reached. This should be done real-time to save default calibration information. Corrections are possible after disconnecting from the O2k. For calibration, follow steps (1) to (5):
=== Air calibration in DatLab ===


1. Select a graph by a click (left mouse button) into the graph or directly by step 2.
::: '''1. Select Graph layout:'''


2. Select the oxygen signal as the active plot by a click on Y1 in the figure legend on the right of the graph. The active plot is highlighted.
:::: Go to menu <span style="background:lightgrey">Layout</span>, check <span style="background:lightgrey">O2</span> and from <span style="background:lightgrey">Standard layouts</span> select <span style="background:lightgrey">01 Calibration show Temp</span>.
This is typically the first layout used after switching on the O2k. Oxygen concentration (blue plots, left Y-axis) and O2 slope (not corrected for instrumental background; red plots, right Y-axis) are displayed on the top graph (left O2k-chamber), and middle graph (right O2k-chamber). The third graph (bottom) shows the block temperature on the left Y-axis and the Peltier power on the right Y-axis. Only when both temperature and Peltier power are constant, the chambers have reached thermal equilibrium. The next step is to observe equilibration of the oxygen signal with a defined gas phase above the stirred aqueous phase ('open' chamber; usually with air as the first step) to perform an oxygen calibration.


3. Only if ‘Mouse Control: Zoom’ mode has been activated: Select “Mouse Control: Mark” in the Graph menu or press Ctrl+M.
::: '''2. Mark'''


4. Set a mark: Hold Shift and click the left mouse button, move the cursor along the time axis, release the left mouse button at the end of the section to be marked.  
:::: Mark a section of the experiment at air saturation, when signal stability is reached. This should be done in real-time to save default calibration information. Corrections are possible after disconnecting from the O2k. For calibration, follow steps (1) to (5):
Remove a section of the mark or the total mark: Move the cursor with Shift+right-click along the time axis, release the mouse button at the end of the section of the mark to be deleted.


5. Rename the mark: Left mouse click on the bar of the mark. Rename the mark for air calibration as “R0”.
:::# Select a graph by a click (left mouse button) into the graph or directly by step 2.
:::# Select the oxygen signal as the active plot by a click on Y1 in the figure legend on the right of the graph. The active plot is highlighted.
:::# Only if <span style="background:lightgrey">Mouse Control: Zoom</span style> mode has been activated: Select <span style="background:lightgrey">Mouse Control: Mark</span style> in the Graph menu or press [Ctrl+M].
:::# Set a mark: Hold [Shift] and click the left mouse button, move the cursor along the time axis, release the left mouse button at the end of the section to be marked. Remove a section of the mark or the total mark: Move the cursor with [Shift+right-click] along the time axis, release the mouse button at the end of the section of the mark to be deleted.
:::# Rename the mark: Left mouse click on the bar of the mark. Rename the mark for air calibration as “R1” (and the mark for zero calibration as “R0”).


'''2. Calibrate'''
::: '''3. Calibrate'''
[[File:Calibration window.PNG|thumb|right|600px|'''Calibration window''']]


1. Open the DatLab calibration window by double clicking on O2 calib. for O2k-chamber A or B in the O2k-status line (bottom left and right). Alternatively, press F5 to open the calibration window for the active plot.
:::# Open the DatLab calibration window by double clicking on <span style="background:yellow">O2 calib.</span style> for O2k-chamber A or B in the O2k-status line (bottom left and right). Alternatively, press [F5] to open the calibration window for the active plot.
:::# Left click on the pull-down button and select the appropriate mark <span style="background:lightgrey">R1</span style>. The average voltage (Raw signal [V]) recorded over the marked section is shown in the corresponding field on the right. The corresponding signal stability is displayed as the uncorrected negative slope of the signal during calibration in [pmol∙s<sup>−1</sup>∙mL<sup>−1</sup>]. Temperature and barometric pressure are displayed as measured over the marked section. Calibration values R1 and R0 can be edited numerically, without exerting an influence on c1. If the temperature or barometric pressure are edited, then c1 is recalculated for the changed conditions.
:::# If not set previously, enter the oxygen solubility factor of the medium, ''F''<sub>M</sub>, relative to pure water. For more info on this factor: [[MiPNet06.03 POS-calibration-SOP]]. For documentation purposes, enter the name of the experimental incubation medium in the corresponding field.
:::# Left click on <span style="background:lightgrey">Calibrate and copy to clipboard</span style>. After clicking on this button all changes in the entire calibration window are applied. The entire plot of oxygen concentration is re-calibrated [µM = nmol/mL], and the corresponding negative slope or volume-specific oxygen flux [pmol∙s<sup>−1</sup>∙mL<sup>−1</sup>] is now based on this new calibration. Click on <span style="background:lightgrey">Calibrate and copy to clipboard</span> before closing the calibration window. Thereby, calibration is activated and calibration parameters are automatically copied to clipboard for entry into the spreadsheet "Template O2 calibration.xlsx".


2. Left click on the pull down button and select the appropriate mark (R0). Many times the zero calibration value is used from a previous experiment. The displayed temperature and pressure are without influence on the calibration calculations for zero oxygen. The stable zero signal, R0, should be <2% of the signal at air saturation, but <5% is acceptable. Most importantly, the zero signal must be stable. The average voltage (Raw signal [V]) recorded over the marked section is shown in the corresponding field on the right. The corresponding signal stability is displayed as the uncorrected negative slope of the signal during calibration in [pmol/(s­.ml)]. Temperature and barometric pressure are displayed as measured over the marked section.
=== Zero calibration in DatLab ===
Calibration values R1 and R0 can be edited numerically, without exerting an influence on c1. If temperature or barometric pressure are edited, then c1 is recalculated for the changed conditions.


3. If not set previously, enter the oxygen solubility factor of the medium, FM, relative to pure water. For more info on this factor: MiPNet06.03 POS-calibration-SOP. For documentation purposes, enter the name of the experimental incubation medium in the coresponding field.
::: '''1. Mark'''


4. Left click on "Calibrate and copy to clipboard". After clicking on this button all changes in the entire calibration window are applied. The entire plot of oxygen concentration is re-calibrated [µM = nmol/ml], and the corresponding negative slope or volume-specific oxygen flux [pmol/(s.ml)] is now based on this new calibration. Calibration parameters are automatically copied to clipboard for entry into the spreadsheet O2 calibration.xlsx.
:::: Mark a section of the experiment after full depletion of oxygen, when signal stability is reached. This should be done in real-time to save default calibration information. Corrections are possible after disconnecting from the O2k. For calibration, follow steps (1) to (5):


:::# Select a graph by a click (left mouse button) into the graph or directly by step 2.
:::# Select the oxygen signal as the active plot by a click on Y1 in the figure legend on the right of the graph. The active plot is highlighted.
:::# Only if <span style="background:lightgrey">Mouse Control: Zoom</span style> mode has been activated: Select <span style="background:lightgrey">Mouse Control: Mark</span style> in the Graph menu or press [Ctrl+M].
:::# Set a mark: Hold [Shift] and click the left mouse button, move the cursor along the time axis, release the left mouse button at the end of the section to be marked. Remove a section of the mark or the total mark: Move the cursor with [Shift+right-click] along the time axis, release the mouse button at the end of the section of the mark to be deleted.
:::# Rename the mark: Left mouse click on the bar of the mark. Rename the mark for zero calibration as “R0”.


== DatLab calibration: real-time vs. disconnected ==
::: '''2. Calibrate'''


DatLab uses calibration values applied real-time (connected to the O2k, recording data) as default values for future experiments. When calibration values are edited in the disconnected mode, they apply only to the current file and will not be used as a new default in experiments. This allows to re-calibrate old files without overwriting the current default values for calibration. Ideally, calibration values that should be used as new defaults are applied real-time when the experiment is still running. However, if the DatLab-calibration is performed after disconnecting, these calibration parameters can be read into other DatLab files using the "Copy from file" function and "Calibrate and copy to clipboard".
:::# Open the DatLab calibration window by double clicking on <span style="background:lightgreen">O2 calib.</span style> for O2k-chamber A or B in the O2k-status line (bottom left and right). Alternatively, press [F5] to open the calibration window for the active plot.
:::# Left click on the pull-down button and select the appropriate mark <span style="background:lightgrey">R0</span style>. Many times the zero calibration value is used from a previous experiment. The displayed temperature and pressure are without influence on the calibration calculations for zero oxygen. The stable zero signal, R0, should be <2 % of the signal at air saturation, but <5 % is acceptable. Most importantly, the zero signal must be stable. The average voltage (Raw signal [V]) recorded over the marked section is shown in the corresponding field on the right. The corresponding signal stability is displayed as the uncorrected negative slope of the signal during calibration in [pmol∙s<sup>−1</sup>∙mL<sup>−1</sup>]. Temperature and barometric pressure are displayed as measured over the marked section. Calibration values R1 and R0 can be edited numerically, without exerting an influence on c1. If the temperature or barometric pressure are edited, then c1 is recalculated for the changed conditions.
:::# If not set previously, enter the oxygen solubility factor of the medium, ''F''<sub>M</sub>, relative to pure water. For more info on this factor: [[MiPNet06.03 POS-calibration-SOP]]. For documentation purposes, enter the name of the experimental incubation medium in the corresponding field.
:::# Left click on <span style="background:lightgrey">Calibrate and copy to clipboard</span style>. After clicking on this button all changes in the entire calibration window are applied. The entire plot of oxygen concentration is re-calibrated [µM = nmol/mL], and the corresponding negative slope or volume-specific oxygen flux [pmol∙s<sup>−1</sup>∙mL<sup>−1</sup>] is now based on this new calibration. Calibration parameters are automatically copied to clipboard for entry into the spreadsheet "Template O2 calibration.xlsx".


Before disconnecting the Oxygraph-2k from DatLab, calibration information is automatically saved and available upon connecting the Oxygraph-2k, even if you exit DatLab and start the program again. The current calibration parameters are displayed when opening the O2 calibration window F5.


=== DatLab calibration: real-time vs. disconnected ===


== Calibration and quality control ==
:::: DatLab uses calibration values applied in real-time (connected to the O2k, recording data) as default values for future experiments. When calibration values are edited in the disconnected mode, they apply only to the current file and will not be used as a new default in experiments. This allows to re-calibrate old files without overwriting the current default values for calibration. Ideally, calibration values that should be used as new defaults are applied in real-time when the experiment is still running. If the DatLab-calibration is performed after disconnecting, these calibration parameters can be read into other DatLab files using the "Copy from file" function and "Calibrate and copy to clipboard".


'''The POS sensor test'''
:::: Before disconnecting the O2k from DatLab, calibration information is automatically saved and available upon re-connecting the O2k, even if you exit DatLab and start the program again. The current calibration parameters are displayed when opening the O2 calibration window [F5].


[[File:Stirrer test.png|thumb|right|500px|'''Stirrer test''' for quality control (standard 30 s) with 30 min time scale displayed with Graph Layout “02-Calibration - Background” (MiR05; 37 °C; data recording interval: 2 s; slope smooting: 40 data points).]]
1. About 20 min are required for approximate air equilibration after temperature equilibration of the incubation medium.


2. Even before final equilibration, perform a [[Stirrer_test|stirrer test]] [F9], switching both stirrers automatically off for 30 s.
=== Calibration and quality control ===


Quality control a: Upon automatic re-start of the stirrer (On), the increase of the oxygen signal should be rapid and monoexponential (see Fig. right, label <span style="background-color:#FFFF00; border-color:#D4D0D0; border-style: solid; border-width: 1px; margin: 2px">'''a'''</span>)
{{Template:Oxygen sensor test}}
 
Quality control b: The raw signal (blue plot; 1 V = 1 µA at gain 1) should be close to 1 to 3 V at 25 to 37 °C at sea level up to 1000 m (pb 101 to 90 kPa). At gain setting of 2 the raw signal [V] is multiplied by 2.
[[File:O2 calibration air.png|thumb|right|500px|'''Quality control''']]
3. Within 40 min, the oxygen signals should be stable with O2 slope (uncorrected) close to zero.
 
Quality control label <span style="background-color:#FFFF00; border-color:#D4D0D0; border-style: solid; border-width: 1px; margin: 2px">'''c'''</span>: Signal noise should be low, reflected in a noise of the O2 slope (red plot) within ±2 (±4 is acceptable) pmol/(s∙ml) at a data recording interval of 2 s and 40 data points selected for calculation of the slope.
 
4. Set a mark on the oxygen signal (R1) and click on O2 Calib. to open the DatLab O2 calibration.
Quality control label <span style="background-color:#FFFF00; border-color:#D4D0D0; border-style: solid; border-width: 1px; margin: 2px">'''d'''</span>: The slope uncorrected should be within ±1 pmol/s∙ml) averaged across the section of the experiment marked as R1 for air calibration (d). The recorded POS signal should be close to the previous calibration under identical experimental conditions. See O2-Calibration window (see Fig. right, label label <span style="background-color:#FFFF00; border-color:#D4D0D0; border-style: solid; border-width: 1px; margin: 2px">'''b’'''</span>).
 
5. Continue with an instrumental O2 background test and perform a zero oxygen calibration.
 
Quality control label label <span style="background-color:#FFFF00; border-color:#D4D0D0; border-style: solid; border-width: 1px; margin: 2px">'''e'''</span>: The zero signal at mark R0 for zero calibration should be <2% of R1 (stable at <5% is acceptable).


== O2-sensor test: when? ==
== O2-sensor test: when? ==


An O2-sensor test should be performed:
::: An O2-sensor test should be performed:
 
1. After switching on the O2k, every day: air calibration, stirrer test, quality control ([[O2_calibration_-_DatLab#Calibration_and_quality_control |Calibration and quality control]]).


2. Zero oxygen calibration: from time to time over weeks; bracketing zero oxygen calibrations when working at low oxygen ([[O2_calibration_-_DatLab#Zero_calibration_-_general_description |Zero calibration]]).
:::# After switching on the O2k, every day: air calibration, stirrer test, quality control ([[O2_calibration_-_DatLab#Calibration_and_quality_control |Calibration and quality control]]).
:::# Zero oxygen calibration: from time to time over weeks; bracketing zero oxygen calibrations when working at low oxygen ([[O2_calibration_-_DatLab#Zero_calibration_-_general_description |Zero calibration]]).
:::# After application of a new membrane and O2-sensor service: in some cases, the signal of the OroboPOS improves (higher signal stability, less noise, shorter response time) when the O2k remains switched on overnight (O2k-Chambers filled with 70 % EtOH or H<sub>2</sub>O if the traces will be evaluated in the morning).
:::# For [[O2_calibration_-_DatLab#Calibration_and_quality_control |O2k-Quality Control]] (O2k-QC) of instrumental performance.
:::# Before an O2k-chamber test ([[O2_calibration_-_DatLab#Instrumental_background_oxygen_flux_-_general |Instrumental O2 background test]]).
:::# During troubleshooting procedures, when switching components between the two chambers, a quick sensor test is performed after each step (stirrer test, sensor signal).


3. After application of a new membrane and O2-sensor service: in some cases, the signal of the OroboPOS improves (higher signal stability, less noise, shorter response time), when the O2k remains switched on over night (O2k-Chambers filled with 70% EtOH).
:::* '''Dynamic POS calibration''' = Calibration of the [[MiPNet06.03 POS-calibration-SOP |response time of the POS]].  
 
:::* '''Static POS calibration''' = Two-point calibration of the [[polarographic oxygen sensor]].
4. For [[O2_calibration_-_DatLab#Calibration_and_quality_control |O2k-Quality Control]] (O2k-QC) of instrumental performance.
 
5. Before an O2k-chamber test ([[O2_calibration_-_DatLab#Instrumental_background_oxygen_flux_-_general |Instrumental O2 background test]]).
 
6. During troubleshooting procedures, when switching components between the two chambers, a quick sensor test is performed after each step (stirrer test, sensor signal).




== Instrumental background oxygen flux - general ==
== Instrumental background oxygen flux - general ==


Instrumental background oxygen flux derives from oxygen fluxes into or out of the respirometer chamber that are not directly related to respiratory activity of the sample. These fluxes are caused by backdiffusion into the chamber at low oxygen pressure, oxygen diffusion out of the chamber at elevated oxygen levels, and oxygen consumption by the polarographic oxygen sensor (OroboPOS). Correction for instrumental background oxygen flux is a standard in high-resolution respirometry, automatically performed by DatLab. Background measurements provide a quality control of instrument function. In the Oroboros O2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become significant and require compliance to standard operating procedures described here.
:::: Instrumental background oxygen flux derives from oxygen fluxes into or out of the respirometer chamber that are not directly related to respiratory activity of the sample. These fluxes are caused by back diffusion into the chamber at low oxygen pressure, oxygen diffusion out of the chamber at elevated oxygen levels, and oxygen consumption by the polarographic oxygen sensor (OroboPOS). Correction for instrumental background oxygen flux is a standard in high-resolution respirometry, performed by DatLab. Background measurements provide quality control of instrument function. In the Oroboros O2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become significant and require compliance to standard operating procedures described here.


== Instrumental background oxygen flux in DatLab ==
:::: Instrumental O<sub>2</sub> background tests can be perfomed using the TIP2k or with manual injections. Detailed instructions are provided here: [[MiPNet14.06_Instrumental_O2_background]]
Instrumental O2 background tests can be perfomed using the TIP2k or with manual injections. Detailed instructions are provided here: [[MiPNet14.06_Instrumental_O2_background]]


Following the preparation of necessary solutions as decribed in [[MiPNet14.06_Instrumental_O2_background]] and loading and mounting of the syringes (when the TIP2k is used) the appropriate DatLab protocol should be used: [[DL-Protocols#DL-Protocol_collection]]
:::: Following the preparation of necessary solutions as described in [[MiPNet14.06_Instrumental_O2_background]] and loading and mounting of the syringes (when the TIP2k is used) the appropriate DatLab protocol should be used: [[Run DL-Protocol/Set O2 limit]]




== Static and dynamic calibration ==
== Troubleshooting ==


::: '''Static POS calibration''' = Two-point calibration of the [[polarographic oxygen sensor]].
* '''Customer ID''':[[Polyak_E | Erzsebet Polyak, University of Wisconsin at Madison]]  


:::: DatLab converts the [[raw signal]] to oxygen concentrations  by applying conversion factors determined during a POS calibration comprising an [[air calibration]] and a [[zero calibration]]. For each [[Gain (O2 channel) | gain]] a different set of calibration values is necessary. Frequently, assumed hardware problems turn out to be a simple calibration problem. This can easily be detected by observing the [[Raw signal]].
:::: Question 1: We have different oxygen levels in chamber A and B.


::: '''Dynamic POS calibration''' = Calibration of the [[MiPNet06.03 POS-calibration-SOP |response time of the POS]].  
:::: Question 2: In chamber A after adding dithionite the signal does not go down below 100 µM.


[[File:Different O2 levels and high zero.PNG|center|600px]]


:::: Answer 1: Different initial oxygen levels after switching on the instrument are due to missing calibration. Sensors need to be [[MiPNet06.03_POS-calibration-SOP | air calibrated]] on each experimental day to account for changes in barometric pressure or to accommodate changes in experimental temperature or medium.


:::: Answer 2: The zero voltage in chamber A is at 0.8996 V, which is too high. Zero voltage should not exceed 5 % (usually lower than 2 %) of the voltage obtained at air saturation (R1). You should perform a complete [[MiPNet19.18B_POS-service | POS service]].




-------------------------
== Keywords: Oxygen signal ==
-------------------------
{{Template:Keywords: Oxygen signal}}


<br />
[[File:Expand.png|right|45px |Click to expand or collaps]]
<div class="toccolours mw-collapsible mw-expanded">
:::[[File:Questions.jpg|left|24px |Explanation of terms]] '''» Keywords'''
<div class="mw-collapsible-content">
::: '''Instrument'''
::::» [[MiPNet19.18B POS-service |O2k-Manual: POS-service]]
::::» [[MiPNet19.03 O2k-cleaning and ISS |O2k-cleaning]]
::::» [[OroboPOS]]
::::» [[Polarographic oxygen sensor]]
::::» [[Stirrer test]]
::: '''Concept'''
::::» [[Air calibration]]
::::» [[Concentration]]
::::» [[High-resolution respirometry]]
::::» [[DL-Protocols#Instrumental_O2_DLP |DL-Protocols]]
::::» [[MitoFit Quality Control System]]
::::» [[MiPNet19.18D O2k-Series G and DatLab 6: Calibration |O2k-Series G and DatLab 6: Calibration]]
::::» [[MiPNet06.03 POS-calibration-SOP |POS-calibration]]
::::» [[Oxygen |Oxygen, dioxygen, O<sub>2</sub>]]
::::» [[Oxygen sensor test]]
::::» [[Oxygen solubility factor]]
::::» [[Zero calibration]]
</div>
</div>


{{Labeling
{{Labeling
Line 205: Line 230:
|additional=DatLab, DL7, DL6a7
|additional=DatLab, DL7, DL6a7
}}
}}
[[Category:Publications]]
{{MitoPedia methods
[[Has title::[[Image:MitoPedia-text.jpg|right|70px|link=MitoPedia|MitoPedia]] Online documentation of O2 calibration in DatLab7.]]
|mitopedia method=Respirometry
[[Was published in year::2017-08-23]]
}}
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=DatLab
}}

Revision as of 14:58, 27 April 2021


                  


O2k-Open Support

Oxygen calibration - DatLab



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

O2 calibration is the calibration in DatLab of the oxygen sensor. It is a prerequisite for obtaining accurate measurements of respiration. Accurate calibration of the oxygen sensor depends on (1) equilibration of the incubation medium with air oxygen partial pressure at the temperature defined by the experimenter; (2) zero oxygen calibration; (3) high stability of the POS signal tested for sufficiently long periods of time; (4) linearity of signal output with oxygen pressure in the range between oxygen saturation and zero oxygen pressure; and (5) accurate oxygen solubility for aqueous solutions for the conversion of partial oxygen pressure into oxygen concentration. The standard oxygen calibration procedure is described below for high-resolution respirometry with the calibration routine using instrumental calibration DL-Protocols in DatLab.


Reference: MiPNet06.03 POS-calibration-SOP, MiPNet12.20_O2k-calibration_tutorial

Calibration-menu.png

Oxygen calibration in instrumental and experimental DatLab files

Calibration window steps.png
Instrumental calibration
It is common practice to save instrumental calibration tests in separate DatLab files, which include (1) air calibration and zero oxygen calibration in the same calibration file, or (2) air calibration and zero oxygen calibration in different calibration files. Step 1: Select calibration sections from Marks (R1, R0) while DatLab is connected to the O2k during an instrumental calibration run. Step 2: Click on Calibrate and copy to clipboard in the O2 calibration window (in menu 'Calibration' > 'Oxygen, O2'). Step 3: Click on Save and disconnect in menu File. The calibration file is saved, and DatLab disconnects from the O2k.
Calibration parameters are automatically transferred to experimental DatLab files
Step 4: When a new DatLab file is opened any time later, the latest calibration parameters are automatically applied in the new DatLab file — independent of maintaining the O2k in operation or switching it off, and independent of maintaining DatLab active or re-opening it after exit.
A calibration can be performed and calibration parameters saved during an experimental run
In the specific case of an experiment that involves an aerobic-anoxic transition, zero oxygen calibration is performed within the experimental file, setting a Mark (R0) on the oxygen signal in the anoxic section, and clicking on Calibrate and copy to clipboard in the O2 calibration window (in menu 'Calibration' > 'Oxygen, O2').
Calibration parameters can be selectively imported from any DatLab file
Step 5: In addition, the calibration parameter for air calibration, zero oxygen calibration, or both can be imported from a selected DatLab file. This is particularly important, if a zero calibration is performed after completion and saving an experimental file. Step 6: A single calibration value (R1 or R0) can be edited manually. Take Step 7 to apply the selected calibration parameters.


Instrumental DL-Protocols

DatLab

Instrumental DL-Protocols guide the O2k operator step-by-step through the calibration procedure, adding automatically the event names and mark names set during the calibration run.

DLP O2-calibration air.png
DLP O2-calibration air-J01.png

Air calibration: O2-calibration_air.DLP

  • O2 sensor test and calibration at air saturation: Performed routinely every day before starting measurements.
Event E: Stirrer test*
1 - Perform a stirrer test [F9], even before final equilibration.
  • Both stirrers are switched off automatically for 30 s (default).
  • Re-start of the stirrers is automatic (automatic event: On).
  • The increase of the O2 signal should be rapid and monoexponential.
2 - Proceed with mark R1.
Mark M: R1 - air calibration
1 - After the stirrer test, the signal should stabilize within 40 min.
  • O2 slope neg. should approach zero, at ± 1 pmol∙s−1∙mL−1.
  • Noise: An 'O2 slope neg.' within ± 2 pmol∙s−1∙mL−1 is optimal (± 4 pmol∙s−1∙mL−1 is acceptable)
2 - Set the marks R1 for each chamber on 'Y1: O2 concentration'.
3 - Open the O2 calibration window in menu 'Calibration' > 'Oxygen, O2'.
4 - Select Mark R1 for 'Air calibration'.
  • 'POS signal: Recorded' should be 1 to 3 V at 25 to 37 °C (up to 1000 m above sea level; pb 101 to 90 kPa) at gain = 1.
  • 'O2 slope neg.' should be within the range of -1 to 1 pmol∙s−1∙mL−1.
5 - Calibrate and copy to clipboard.
6 - Paste into the Excel file: "Template O2 calibration.xlsx".
7 - Proceed with event 1C.
Event E: 1C - Close chamber
1 - After air calibration, close the chamber by gently pushing the stopper fully downwards.
2 - Siphon off excess medium from the top of the stopper using the Integrated Suction System (ISS).
3 - Check that any air bubbles are removed from the chamber.
4 - Switch to 'Y2: O2 slope neg.' by clicking on the Y2 label on the graph legend.
5 - Proceed with mark J°1.
Mark M: J°1 - O2 BG flux ~ air saturation (instrumental background flux after air calibration)
1 - Set a mark (J°1) after event 1C at a section, where the O2 slope is stable, and observe the signal.
  • 'O2 slope neg.' should be within the range of ± 2 to 4 pmol∙s−1∙mL−1.
  • 'O2 slope neg.' > 4 pmol∙s−1∙mL−1 may indicate a biological contamination in the medium or O2k-chamber.
2 - Save the file in menu 'File' > 'Save and disconnect' before starting an experiment.
  • Although the calibration file is closed, the values will be automatically used as default calibration values for the next measurements.
DLP O2-calibration air and zero.png
DatLab

Air and zero calibration: O2-calibration_air and zero.DLP

  • O2 sensor test and calibration at air saturation and zero oxygen:
This DL-Protocol is identical in the section of air calibration, and continues with zero calibration as follows:
Mark M: J°1 - O2 BG flux ~ air saturation (instrumental background flux after air calibration)
1 - Set a mark (J°1) after event 1C at a section, where the O2 slope is stable, and observe the signal.
  • 'O2 slope neg.' should be within the range of 2 to 4 pmol∙s−1∙mL−1.
  • 'O2 slope neg.' > 4 pmol∙s−1∙mL−1 may indicate a biological contamination in the medium or O2k-chamber.
2 - Switch to plot 'Y1: O2 concentration' by clicking on the Y1 label on the graph legend.
3 - Proceed with event 2Dith*.
Event E: 2Dith* - Manual dithionite titration
1 - Events 2Dith*: Inject fresh Na-dithionite solution ("zero solution") until the O2 signal is reduced close to 0 µM.
  • A few minutes are required until a stable "zero" signal is obtained
  • Repeat until the signal does not respond by a further decline. Siphon off excess medium from the stopper.
2 - Switch to plot 'Y1: O2 concentration' by clicking on the Y1 label on the graph legend.
3 - Proceed with mark R0.
Mark M: R0 - Zero O2 calibration
1 - Set a mark (R0) after the last event 2Dith* when the signal is stable (Y1: O2 concentration).
2 - Open the O2 calibration window in menu 'Calibration' > 'Oxygen, O2'.
3 - Select mark R0 for 'Zero calibration'.
  • R0 should be stable and < 2% of the signal at air saturation, but < 5% is acceptable.
4 - Select mark R1 for 'Air calibration'.
5 - Calibrate and copy to clipboard.
6 - Paste into the Excel file: "Template O2 calibration.xlsx".
7 - Save the file in menu 'File' > 'Save and disconnect' before starting the next file.
  • After the calibration file is saved while connected to the O2k, the calibration parameters will be automatically used as default after restarting DatLab.


Oxygen calibration - general description

The general description lists the calibration steps in further detail, without considering the convenience of using the instrumental DL-Protocols listed above.
DatLab converts the oxygen signal of the polarographic oxygen sensor (OroborPOS or POS) to oxygen concentration during calibration comprising an air calibration (R1) and a zero calibration (R0).
  1. Appropriate calibration conditions are obtained in the O2k-chamber at experimental temperature.
  2. R1 and R0 are marked on the (blue) plot of oxygen concentration.
  3. The marks for R1 and R0 are pulled down in the O2 calibration window of DatLab.
  4. Click on Calibrate and copy to clipboard before closing the calibration window.
Frequently, assumed hardware problems turn out to be a simple calibration problem.
Air saturation is achieved by stirring the aqueous medium in contact with air in the O2k-chamber without sample.
Stopper-spacers used for air calibration with white PVDF stopper (Chamber A) or black PEEK stopper (Chamber B).
  1. Add incubation medium into the chambers with an excess volume of at least 0.1 mL above the experimental chamber volume (2 mL or 0.5 mL) in order to fill the O2k-chamber and injection capillary of the stopper when it is fully inserted (closed). The volume does not have to be accurate, as long as it is above the minimum volume. Switch on the stirrers either during or after the addition of the medium.
  2. Insert the stoppers slowly to their volume-calibrated position. Suck off excess medium ejected through the injection capillary and remaining in the well of the stopper. Then lift the stoppers using the stopper-spacer tool, leaving a gas volume above the liquid phase for final air equilibration - Open chamber position. The central level of the gas phase remains above the rotating stirrer bar, preventing bubbles and foam from being formed which would block gas exchange. To ensure a well-defined pO2 in the gas phase, the gas volume has to be renewed (exchanged for air), if the medium was originally not near air saturation. This is achieved simply by fully inserting and re-opening the stopper. Equilibration is a slow process: stability should be reached within one hour (figure to the right). A stirrer test [F9] is performed early during equilibration.
  3. After stabilization of the POS signal, the recorded signal at air saturation, R1, is about 1-3 V (at Gain 1) and a temperature of 25-37 °C. A signal of 1 V corresponds to a signal current of the POS of 1 µA (corresponding to 2 V at Gain 2). Under all experimental conditions, the raw signal must be <10 V. Continue recording for 3-10 min to check for signal stability. You may proceed at this point with an O2 background test (see below).
Air calibration: Oxygen concentration (blue plot; 180 µM; scale from 0 to 250 nmol/mL or 0 to 250 µM) over 1 h after switching on the O2k (Power-O2k P4, chamber A) and setting the experimental temperature at 37 °C, using medium stirred for equilibration with a gas phase of air at 575 m altitude. The red plot is the negative slope of oxygen concentration over time [pmol∙s−1∙mL−1] on the right Y-axis, with zero in the middle position. A slope of zero (for uncorrected ‘O2 Slope neg.’) indicates a constant O2 signal over time. (2014-07-24 P4-01.DLD)
Zero oxygen calibration may be achieved by chemical or biological depletion of oxygen. (1) Chemical: Following chamber assembly or exchange of the POS membrane, zero solution (Na-dithionite, OroboPOS-Service Kit) is titrated into the chamber, which is part of the automatic TIP2k-supported instrumental background test (MiPNet14.06). (2) Biological: Zero calibration can be performed with mitochondria or cell suspensions that allow complete oxygen depletion.

Air calibration in DatLab

1. Select Graph layout:
Go to menu Layout, check O2 and from Standard layouts select 01 Calibration show Temp.

This is typically the first layout used after switching on the O2k. Oxygen concentration (blue plots, left Y-axis) and O2 slope (not corrected for instrumental background; red plots, right Y-axis) are displayed on the top graph (left O2k-chamber), and middle graph (right O2k-chamber). The third graph (bottom) shows the block temperature on the left Y-axis and the Peltier power on the right Y-axis. Only when both temperature and Peltier power are constant, the chambers have reached thermal equilibrium. The next step is to observe equilibration of the oxygen signal with a defined gas phase above the stirred aqueous phase ('open' chamber; usually with air as the first step) to perform an oxygen calibration.

2. Mark
Mark a section of the experiment at air saturation, when signal stability is reached. This should be done in real-time to save default calibration information. Corrections are possible after disconnecting from the O2k. For calibration, follow steps (1) to (5):
  1. Select a graph by a click (left mouse button) into the graph or directly by step 2.
  2. Select the oxygen signal as the active plot by a click on Y1 in the figure legend on the right of the graph. The active plot is highlighted.
  3. Only if Mouse Control: Zoom mode has been activated: Select Mouse Control: Mark in the Graph menu or press [Ctrl+M].
  4. Set a mark: Hold [Shift] and click the left mouse button, move the cursor along the time axis, release the left mouse button at the end of the section to be marked. Remove a section of the mark or the total mark: Move the cursor with [Shift+right-click] along the time axis, release the mouse button at the end of the section of the mark to be deleted.
  5. Rename the mark: Left mouse click on the bar of the mark. Rename the mark for air calibration as “R1” (and the mark for zero calibration as “R0”).
3. Calibrate
Calibration window
  1. Open the DatLab calibration window by double clicking on O2 calib. for O2k-chamber A or B in the O2k-status line (bottom left and right). Alternatively, press [F5] to open the calibration window for the active plot.
  2. Left click on the pull-down button and select the appropriate mark R1. The average voltage (Raw signal [V]) recorded over the marked section is shown in the corresponding field on the right. The corresponding signal stability is displayed as the uncorrected negative slope of the signal during calibration in [pmol∙s−1∙mL−1]. Temperature and barometric pressure are displayed as measured over the marked section. Calibration values R1 and R0 can be edited numerically, without exerting an influence on c1. If the temperature or barometric pressure are edited, then c1 is recalculated for the changed conditions.
  3. If not set previously, enter the oxygen solubility factor of the medium, FM, relative to pure water. For more info on this factor: MiPNet06.03 POS-calibration-SOP. For documentation purposes, enter the name of the experimental incubation medium in the corresponding field.
  4. Left click on Calibrate and copy to clipboard. After clicking on this button all changes in the entire calibration window are applied. The entire plot of oxygen concentration is re-calibrated [µM = nmol/mL], and the corresponding negative slope or volume-specific oxygen flux [pmol∙s−1∙mL−1] is now based on this new calibration. Click on Calibrate and copy to clipboard before closing the calibration window. Thereby, calibration is activated and calibration parameters are automatically copied to clipboard for entry into the spreadsheet "Template O2 calibration.xlsx".

Zero calibration in DatLab

1. Mark
Mark a section of the experiment after full depletion of oxygen, when signal stability is reached. This should be done in real-time to save default calibration information. Corrections are possible after disconnecting from the O2k. For calibration, follow steps (1) to (5):
  1. Select a graph by a click (left mouse button) into the graph or directly by step 2.
  2. Select the oxygen signal as the active plot by a click on Y1 in the figure legend on the right of the graph. The active plot is highlighted.
  3. Only if Mouse Control: Zoom mode has been activated: Select Mouse Control: Mark in the Graph menu or press [Ctrl+M].
  4. Set a mark: Hold [Shift] and click the left mouse button, move the cursor along the time axis, release the left mouse button at the end of the section to be marked. Remove a section of the mark or the total mark: Move the cursor with [Shift+right-click] along the time axis, release the mouse button at the end of the section of the mark to be deleted.
  5. Rename the mark: Left mouse click on the bar of the mark. Rename the mark for zero calibration as “R0”.
2. Calibrate
  1. Open the DatLab calibration window by double clicking on O2 calib. for O2k-chamber A or B in the O2k-status line (bottom left and right). Alternatively, press [F5] to open the calibration window for the active plot.
  2. Left click on the pull-down button and select the appropriate mark R0. Many times the zero calibration value is used from a previous experiment. The displayed temperature and pressure are without influence on the calibration calculations for zero oxygen. The stable zero signal, R0, should be <2 % of the signal at air saturation, but <5 % is acceptable. Most importantly, the zero signal must be stable. The average voltage (Raw signal [V]) recorded over the marked section is shown in the corresponding field on the right. The corresponding signal stability is displayed as the uncorrected negative slope of the signal during calibration in [pmol∙s−1∙mL−1]. Temperature and barometric pressure are displayed as measured over the marked section. Calibration values R1 and R0 can be edited numerically, without exerting an influence on c1. If the temperature or barometric pressure are edited, then c1 is recalculated for the changed conditions.
  3. If not set previously, enter the oxygen solubility factor of the medium, FM, relative to pure water. For more info on this factor: MiPNet06.03 POS-calibration-SOP. For documentation purposes, enter the name of the experimental incubation medium in the corresponding field.
  4. Left click on Calibrate and copy to clipboard. After clicking on this button all changes in the entire calibration window are applied. The entire plot of oxygen concentration is re-calibrated [µM = nmol/mL], and the corresponding negative slope or volume-specific oxygen flux [pmol∙s−1∙mL−1] is now based on this new calibration. Calibration parameters are automatically copied to clipboard for entry into the spreadsheet "Template O2 calibration.xlsx".


DatLab calibration: real-time vs. disconnected

DatLab uses calibration values applied in real-time (connected to the O2k, recording data) as default values for future experiments. When calibration values are edited in the disconnected mode, they apply only to the current file and will not be used as a new default in experiments. This allows to re-calibrate old files without overwriting the current default values for calibration. Ideally, calibration values that should be used as new defaults are applied in real-time when the experiment is still running. If the DatLab-calibration is performed after disconnecting, these calibration parameters can be read into other DatLab files using the "Copy from file" function and "Calibrate and copy to clipboard".
Before disconnecting the O2k from DatLab, calibration information is automatically saved and available upon re-connecting the O2k, even if you exit DatLab and start the program again. The current calibration parameters are displayed when opening the O2 calibration window [F5].


Calibration and quality control

Stirrer test for quality control (standard 30 s) with 30 min time scale displayed with Graph Layout “02-Calibration - Background” (MiR05; 37 °C; data recording interval: 2 s; slope smoothing: 40 data points).
Oxygen sensor test
The oxygen sensor test starts in an open chamber.
1. Before final equilibration, perform a stirrer test [F9], switching both stirrers automatically off for 30 s.
2. About 20 minutes are required for air equilibration after temperature equilibration of the incubation medium, visualized as stabilization of the Peltier power (Fig. Quality control; time scale is 01:10 hh:mm).
  • Quality control label a: Upon automatic re-start of the stirrer (On), the increase of the oxygen signal should be rapid and monoexponential.
  • Quality control label b: The raw signal (blue plot; 1 V = 1 µA at Gain 1) should be 1 to 3 V at 25 to 37 °C at sea level up to 1000 m (pb 101 to 90 kPa). At a Gain setting of 2 the raw signal [V] is multiplied by 2.
Quality control
3. Within 40 minutes, the oxygen signals should be stable with an O2 slope (uncorrected) close to zero.
  • Quality control label c: Signal noise should be low, reflected in a noise of the O2 slope (red plot) within ± 2 (± 4 is acceptable) pmol∙s−1∙mL−1 at a data recording interval of 2 s and 40 data points selected for calculation of the slope.
4. Set a mark on the oxygen signal (R1) and click on O2 Calib. to open the DatLab O2 calibration window (MiPNet06.03 POS-calibration-SOP).
  • Quality control label d: The slope uncorrected should be within ± 1 pmol∙s−1∙mL−1 averaged across the section of the experiment marked as R1 for air calibration (d). The recorded POS signal should be close to the previous calibration under identical experimental conditions. See O2-Calibration window (see Fig. right, label b’).
5. Close the chamber and if required, perform a zero oxygen calibration.
  • Quality control label e: After closing the chamber, select plot Y2 and set mark J°1. O2 slope neg. should be within 3.0 ± 1 pmol∙s−1∙mL−1.
O2 slope neg. values higher than 4.0 pmol∙s−1∙mL−1 indicate:
» Biological contamination.
» Air bubbles in the Closed chamber: switch on the illumination of the O2k and inspect the O2k-Chamber through the front window. Remove any air bubbles.
» A large volume of medium collected in the receptacle of the stopper: siphon off excess medium.
» A larger chamber volume: check O2k-Chamber volume calibration.
  • Quality control label f: The zero signal at mark R0 for zero calibration should be <2 % of R1 (stable at <5 % is acceptable).

O2-sensor test: when?

An O2-sensor test should be performed:
  1. After switching on the O2k, every day: air calibration, stirrer test, quality control (Calibration and quality control).
  2. Zero oxygen calibration: from time to time over weeks; bracketing zero oxygen calibrations when working at low oxygen (Zero calibration).
  3. After application of a new membrane and O2-sensor service: in some cases, the signal of the OroboPOS improves (higher signal stability, less noise, shorter response time) when the O2k remains switched on overnight (O2k-Chambers filled with 70 % EtOH or H2O if the traces will be evaluated in the morning).
  4. For O2k-Quality Control (O2k-QC) of instrumental performance.
  5. Before an O2k-chamber test (Instrumental O2 background test).
  6. During troubleshooting procedures, when switching components between the two chambers, a quick sensor test is performed after each step (stirrer test, sensor signal).


Instrumental background oxygen flux - general

Instrumental background oxygen flux derives from oxygen fluxes into or out of the respirometer chamber that are not directly related to respiratory activity of the sample. These fluxes are caused by back diffusion into the chamber at low oxygen pressure, oxygen diffusion out of the chamber at elevated oxygen levels, and oxygen consumption by the polarographic oxygen sensor (OroboPOS). Correction for instrumental background oxygen flux is a standard in high-resolution respirometry, performed by DatLab. Background measurements provide quality control of instrument function. In the Oroboros O2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become significant and require compliance to standard operating procedures described here.
Instrumental O2 background tests can be perfomed using the TIP2k or with manual injections. Detailed instructions are provided here: MiPNet14.06_Instrumental_O2_background
Following the preparation of necessary solutions as described in MiPNet14.06_Instrumental_O2_background and loading and mounting of the syringes (when the TIP2k is used) the appropriate DatLab protocol should be used: Run DL-Protocol/Set O2 limit


Troubleshooting

* Customer ID: Erzsebet Polyak, University of Wisconsin at Madison 
Question 1: We have different oxygen levels in chamber A and B.
Question 2: In chamber A after adding dithionite the signal does not go down below 100 µM.
Different O2 levels and high zero.PNG
Answer 1: Different initial oxygen levels after switching on the instrument are due to missing calibration. Sensors need to be air calibrated on each experimental day to account for changes in barometric pressure or to accommodate changes in experimental temperature or medium.
Answer 2: The zero voltage in chamber A is at 0.8996 V, which is too high. Zero voltage should not exceed 5 % (usually lower than 2 %) of the voltage obtained at air saturation (R1). You should perform a complete POS service.


Keywords: Oxygen signal


Questions.jpg


Click to expand or collaps



Labels: MiParea: Instruments;methods 





HRR: Oxygraph-2k, O2k-Fluorometer, O2k-Protocol 

DatLab, DL7, DL6a7 


MitoPedia methods: Respirometry 


MitoPedia O2k and high-resolution respirometry: DatLab