Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

SUIT-026 O2 mt D063

From Bioblast
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.


high-resolution terminology - matching measurements at high-resolution


SUIT-026 O2 mt D063

Description

1S;2Rot;3D;3c;4Ama.png

Abbreviation: RET (respiratory control) of SUIT-026 AmR mt D064

Reference: B - SUIT-026 - Protocol for the respiratory control of SUIT-026 AmR mt D064

SUIT number: D063_1S;2Rot;3D;3c;4Ama

O2k-Application: O2


SUIT-026 O2 mt D063 has been designed to serve as a respiratory control of the SUIT-026 AmR mt D064 (protocol to study the RET-initiated ROS production in mitochondrial preparations (isolated mitochondria, tissue homogenate and permeabilized cells which are already permeabilized when they are added to the chamber).

Communicated by  Komlodi T, Gnaiger E (last update 2019-09-24)


Representative traces

SUIT-026 O2 mt D063 O2 trace.png

MitoPedia: SUIT

Steps and respiratory states

1S;2Rot;3D;4Ama.png


Step State Pathway Q-junction Comment - Events (E) and Marks (M)
1S SL(n) S CII 1S
2Rot SL(n) S CII 1S;2Rot
3D SP S CII 1S;2Rot;3D
3c SP S CII 1S;2Rot;3D;3c
4Ama ROX 1S;2Rot;3D;3c;4Ama
  • Rox is the residual oxygen consumption in the ROX state, due to oxidative side reactions, estimated after addition of antimycin A (inhibitor of CIII). Rox is subtracted from oxygen flux as a baseline for all respiratory states, to obtain mitochondrial respiration (mt).


Questions.jpg


Click to expand or collaps

Strengths and limitations

  • This protocol serves as a respiratory control (since fluorescence dyes may inhibit components of the ET system) and quality control for the sample for SUIT-026 AmR mt D064.
+ Short protocol.
+ The addition of cytochrome c to assess the mt outer membrane integrity provides a quality control of the sample, which cannot be performed in the protocol with Amplex UltraRed.
- This protocol does not provide information about all the coupling control states (LEAK, OXPHOS and ET). However, it is possible to create a DLPU by additing an Event&Mark for the uncoupler titration (4U) after ADP (3D) to obtain ET-state.


Compare SUIT protocols


Chemicals and syringes

Step Chemical(s) and link(s) Comments
1S Succinate (S)
2Rot Rotenone (Rot)
3D ADP (D)
3c Cytochrome c (c)
4Ama Antimycin A (Ama)
Suggested stock concentrations are shown in the specific DL-Protocol.


References

MitoPedia concepts: SUIT protocol, SUIT B, Find 


MitoPedia methods: Fluorometry