Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

SUIT reference protocol

From Bioblast
Revision as of 07:27, 9 September 2016 by Gnaiger Erich (talk | contribs)


high-resolution terminology - matching measurements at high-resolution


SUIT reference protocol

Description

The substrate-uncoupler-inhibitor titration (SUIT) reference protocol, SUIT RP, provides a common baseline for comparison of mitochondrial respiratory control in a large variety of species, tissues and cell types, mt-preparations and laboratories, for establishing a database on comparative mitochondrial phyisology. The SUIT RP consists of two harmonized SUIT protocols (SUIT_FNSGp(PGM)01 - SUIT RP1 and SUIT_FNSGp(PGM)02 - SUIT RP2). These are coordinated such that they can be statistically evaluated as replicate measurements of cross-linked respiratory states, while additional information is obtained when the two protocols are conducted in parallel. Therefore, these harmonized SUIT protocols are complementary with their focus on specific respiratory coupling and pathway control aspects, extending previous strategies for respirometrc OXPHOS analysis.

SUIT_RP1: SUIT_FNSGp(PGM)01 - FNSGp(PGM)01&Dc-CIV: FNSGp 1PM 2D&Dc 3U 4G 5S 6Oct 7Rot 8Gp 9Ama 10Tm 11Azd
SUIT_RP2: SUIT_FNSGp(PGM)02 - FNSGp(PGM)02&Mc-CIV 1D 2Oct 3M&Mc 4P 5G 6S 7Gp 8U 9Rot 10Ama 11Tm 12Azd

Abbreviation: SUIT RP

Reference: MiPNet21.06 SUIT RP


MitoPedia concepts: SUIT concept 

SUIT-catg

Cross-linked respiratory states

The cross-linked respiratory states for harmonization between protocols are:

SUIT RP