Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Stirrer test"

From Bioblast
Line 1: Line 1:
{{MitoPedia
{{MitoPedia
|abbr=F9
|abbr=F9
|description=A '''stirrer test''' is performed in the [[O2k]] for quick evaluation of the performance of the [[OroboPOS]] and for [[POS calibration - dynamic|dynamic calibration]]. The stirrer test is initiated by pressing [F9] in [[DatLab]]. Stirring is stopped in both chambers and restarted after a selected period. The default period is 30 s, for experiments at 37 °C. At lower experimental temperature, this period should be prolonged (60 s at 25 °C). In the [[O2k-technical_support_and_open_innovation#O2k_Quality_Control |SOP (I2k Quality Control) for the O<sub>2</sub> sensor test]], the stirrer test is performed in the 'open' chamber in oonjunction with air calibration. In general, the stirrer test can be performed equally with an open or closed chamber.  
|description=A '''stirrer test''' is performed in the [[OROBOROS O2k]] for quick evaluation of the performance of the [[OroboPOS]] and for [[POS calibration - dynamic|dynamic calibration]]. The stirrer test is initiated by pressing [F9] in [[DatLab]]. Stirring is stopped in both chambers and restarted after a selected period. The default period is 30 s, for experiments at 37 °C. At lower experimental temperature, this period should be prolonged (60 s at 25 °C). In the [[O2k-technical_support_and_open_innovation#O2k_Quality_Control |SOP (I2k Quality Control) for the O<sub>2</sub> sensor test]], the stirrer test is performed in the 'open' chamber in oonjunction with air calibration. In general, the stirrer test can be performed equally with an open or closed chamber.  
|info=[[MiPNet06.03 POS-calibration-SOP]]
|info=[[MiPNet06.03 POS-calibration-SOP]]
}}
}}

Revision as of 09:04, 6 July 2017


high-resolution terminology - matching measurements at high-resolution


Stirrer test

Description

A stirrer test is performed in the OROBOROS O2k for quick evaluation of the performance of the OroboPOS and for dynamic calibration. The stirrer test is initiated by pressing [F9] in DatLab. Stirring is stopped in both chambers and restarted after a selected period. The default period is 30 s, for experiments at 37 °C. At lower experimental temperature, this period should be prolonged (60 s at 25 °C). In the SOP (I2k Quality Control) for the O2 sensor test, the stirrer test is performed in the 'open' chamber in oonjunction with air calibration. In general, the stirrer test can be performed equally with an open or closed chamber.

Abbreviation: F9

Reference: MiPNet06.03 POS-calibration-SOP


MitoPedia O2k and high-resolution respirometry: DatLab