| == Troubleshooting ==
| |
|
| |
|
| === '''Oxygen slope neg. fluctuating'''ย ===
| |
| ::::* '''Customer ID''': Anonymous
| |
| ::::'''Question:'''ย I am having some trouble getting the O<sub>2</sub> slope neg. to stabilize at zero pmol/(sโขmL) during air calibration. The signal will be stable at -4 pmol/(sโขmL), then fairly quickly climb to +4 pmol/(sโขmL) then swap back down to -4 pmol/(sโขmL) in a long wave that repeats periodically over time.
| |
|
| |
| [[File:Before changing energy supply.png|600px|center]]
| |
| ::::'''Answer:''' We noticed that your peltier power follows the changes observed in the O<sub>2</sub> slope neg. To rule out electrical fluctuations we would ask you to exchange the energy supply of the O2k and run an air calibration file to check.
| |
| [[File:After changing energy supply.png|600px|center|After changing energy supply]]
| |