== Troubleshooting ==


=== '''Oxygen slope neg. fluctuating'''Β  ===
::::* '''Customer ID''': Anonymous
::::'''Question:'''Β  I am having some trouble getting the O<sub>2</sub> slope neg. to stabilize at zero pmol/(sβ€’mL) during air calibration. The signal will be stable at -4 pmol/(sβ€’mL), then fairly quickly climb to +4 pmol/(sβ€’mL) then swap back down to -4 pmol/(sβ€’mL) in a long wave that repeats periodically over time.
[[File:Before changing energy supply.png|600px|center]]
::::'''Answer:''' We noticed that your peltier power follows the changes observed in the O<sub>2</sub> slope neg. To rule out electrical fluctuations we would ask you to exchange the energy supply of the O2k and run an air calibration file to check.
[[File:After changing energy supply.png|600px|center|After changing energy supply]]
:::: After changing the electrical supply the O2 slope neg. and the Peltier Power signal were stable.