'''Question:'''
Previous Product ID 12300-01 (PVDF stopper)
If only Ca+2 is detected with the suitable module? Or, other ions such Na+, phosphate, K+ can also be detected?
Previous Product ID 32000-01


'''Answer:'''
[[Image:BB-Bioblast.jpg|left|40px|link=http://www.bioblast.at/index.php/Bioblast:About|Bioblast wiki]]
We offer membranes to detect TPP+ and Ca2+. There are ion selective membranes available e.g. from Sigma Aldrich for a wide variety of species. If cut to a circle with a diameter of 4 mm these membranes can be used in our ISE system.
== Popular Bioblast page ==
 
:::: [[O2k-TPP+ ISE-Module]] has been accessed more than
Many ions may also be detected by fluorimetric based methods, see [[http://wiki.oroboros.at/index.php/O2k-Fluorescence_LED2-Module]]
::::* 20,000 times (2019-07-22)
 
::::* 15,000 times (2017-01-30)
 
::::* 10,000 times (2016-02-22)
 
::::*  5,000 times (2014-08-13)
 
--[[User:Fasching Mario|Fasching Mario]] 09:48, 22 February 2012 (CET)
 
== Types of connectors ==
 
'''Question:''' Are the ISE sockets a BNC cable connection?
 
'''Answer:''' In Series D and higher the ISE connectors consists of 2 ports: 1.) a BNC port: both poles are "live" so it can be used either for a  combination electrode (like a classical pH) or for just one singe  electrode that requires a separate reference electrode 2.) a 2 mm pin type connection that is used for connecting the separate reference electrodes we use for TPP and now also for pH This 2 mm pin port connects to the same electronics as the outer channel of the BNC port. See also chapter 3.2 of the ISE Manual [http://www.oroboros.at/index.php?id=o2k-multisensor-manual MiPNet15.03]. In the same manual you find the relevant information for older series of the oxygraph.[[User:Fasching Mario|Fasching Mario]] 09:22, 19 September 2012 (CEST)
 
== Long term storage of the TPP electrode ==
 
Question: What is the procedure for long term storage of the TPP electrode?
 
Answer: For long time storage (several months) please disassemble the electrode, wash all parts with ethanol (sterilization), allow all parts to become  completely dry, then re-assemble the dry electrode without a membrane(to  have everything together and to avoid loosing  small parts). Then store  the electrode in a dark place like the TPP accessory box. [[User:Fasching Mario|Fasching Mario]] 14:41, 26 September 2012 (CEST)
 
== Experineces with the Ca2+ electrode ==
 
created [[User:Fasching Mario|Fasching Mario]] 10:16, 7 November 2012 (CET)
 
=== Ca2+ electrode and FCCP ===
 
There are strong indications that FCCP and the [[ISE-Ca2+ Membranes]] are incompatible....[[User:Fasching Mario|Fasching Mario]] 10:19, 7 November 2012 (CET)