{{Technical support}}
{{MitoPedia
|description=Biological contamination may be caused by microbial growth in the O2k-Chamber or in the experimental medium.
}}
__TOC__
{{Technical support integrated}}
:::: Biological contamination of the O2k-Chamber is detected as a high O<sub>2</sub> flux in the [[Closed chamber |'''closed chamber''']], measured in a [[Oxygen sensor test |'''QC1: Oxygen sensor test''']] or [[Oxygen flux - instrumental background |'''QC2: Instrumental O<sub>2</sub> background test''']].


ย 
:::: The problem can be solved by first cleaning the O2k-Chamber by chemical sterilization (70 % ethanol with 30 % water; not 100 % ethanol) and test with a freshly prepared experimental medium, such as [[MiR05]]. Make sure the 70% ethanol does not contain any additives; ''e.g.'', 70 % ethanol used in hospital settings may contain antiseptics ([[Talk:MiPNet19.03 O2k-cleaning and ISS#Problem solved |EtOH containing antiseptics]] provided by [[Garcia-Roves PM]]). If repeated washing and storage with 70 % ethanol does not remove a biological contamination, the glass chamber has to be cleaned as described for protein contamination under [[MiPNet19.03 O2k-cleaning and ISS |O2k-Chamber cleaning with HCl]] by removing the glass chamber from the O2k and using a strong acid.
Biological contaminationย  is caused by the growth of biological material (originating from samples or otherwise). It should not be confused with other kinds of chamber contamination mentioned in [[Cleaning the glass chamber]].
ย 
Biological contamination of the chamber is typically characterized by a high O2 flux with a closed chamber near air saturation. It is detected by an [[Oxygen sensor test]], an [[instrumental background test]] or by observing the flux at closed chamber for a short time routinely before any experiment. In the latter two cases the high flux contamination may be caused by biological contamination of the chamber itself or by a contaminated medium. This can be tested by observing the flux in pure water.
ย 
The ideal remedial agent is 70% ethanol with 30% water (not 100% ethanol). Make sure the 70% ethanol does not contain any additives. E.g. 70% ethanol used in hospital settings may contain antiseptics (information provided by [[Garcia-Roves PM]]).
Prevention of biological contamination of the chamber is primarily by storage under 70% ethanol. If repeated washing with 70% Ethanol does not remove an existing biological contamination, the glass chamber has to be cleaned as described for protein contamination under [[Cleaning the glass chamber]] by removing the glass chamber from the instrument and using a strong acid.
ย 
ย 
{{#set: Technical service =Chamber}}
[[Category:Technical service]]

Latest revision as of 16:29, 5 November 2020


high-resolution terminology - matching measurements at high-resolution


Biological contamination

Description

Biological contamination may be caused by microbial growth in the O2k-Chamber or in the experimental medium.


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Biological contamination of the O2k-Chamber is detected as a high O2 flux in the closed chamber, measured in a QC1: Oxygen sensor test or QC2: Instrumental O2 background test.
The problem can be solved by first cleaning the O2k-Chamber by chemical sterilization (70 % ethanol with 30 % water; not 100 % ethanol) and test with a freshly prepared experimental medium, such as MiR05. Make sure the 70% ethanol does not contain any additives; e.g., 70 % ethanol used in hospital settings may contain antiseptics (EtOH containing antiseptics provided by Garcia-Roves PM). If repeated washing and storage with 70 % ethanol does not remove a biological contamination, the glass chamber has to be cleaned as described for protein contamination under O2k-Chamber cleaning with HCl by removing the glass chamber from the O2k and using a strong acid.