Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Measuring hydrogen peroxide"

From Bioblast
(Redirected page to Hydrogen peroxide)
Β 
(16 intermediate revisions by 3 users not shown)
Line 1: Line 1:
Two techniques to measure H<sub>2</sub>O<sub>2</sub> concentration and H<sub>2</sub>O</sub>2 production can be combined with the O2k: The [[O2k-Fluorescence Module|fluorescence measurement]] using Amplex Red, and the electrochemical measurement with a H<sub>2</sub>O<sub>2</sub> electrode.
#REDIRECT [[Hydrogen peroxide]]
Β 
== Fluorescence ==
The fluorimetric measurement of H<sub>2</sub>O<sub>2</sub> in the O2k is based on the [[O2k-Fluorescence Module]], using Amplex Red (or one of its brand names).<ref>[[Sumbalova 2011 Abstract Kagoshima|Sumbalova Z, Harrison DK, Gradl P, Fasching M, Gnaiger E (2011) Mitochondrial membrane potential, coupling control, H2O2 production, and the upper limit of mitochondrial performance. Abstract Kagoshima]].</ref> The O2k-Fluorescence Module consists of a LED light source and a photodiode + filter attached to the O2k-glass window.<ref>[[Fasching 2011 Abstract Berlin|Fasching M, Harrison DK, Tretter L, Gnaiger E (2011) Combination of high-resolution respirometry and fluorimetry for continuous monitoring of hydrogen peroxide production by mitochondria with resolution in the nanomolar range. Abstract Berlin]]</ref> The potentiometric channel of the O2k-MultiSensor System is used to directly record the signal in DatLab ([[Hickey 2011 JCPB]]; [[MitoCom K-Regio|''MitoCom'']]. In addition, the Oxygraph-2k has been coupled to full fluorescence spectrophotometers for this purpose, with a light guide inserted through the [[MultiSensor stopper PEEK black conical, central and 2.5 mm side port|black PEEK stopper]] [[Anderson 2010 AJP]].
Β 
== Electrochemical Measurement ==
Β 
For electrochemically (H<sub>2</sub>O<sub>2</sub> electrode) measuring H<sub>2</sub>O<sub>2</sub> (and all other amperometric methods) please consult our" NO" or "amperometric"
[http://www.oroboros.at/index.php?o2k_no-manual manual]. All stated there about measuring NO is technically also correct for using a H<sub>2</sub>O<sub>2</sub> electrode together with the oxygraph. In this manual you will also find which accessories you need to use a H<sub>2</sub>O<sub>2</sub> electrode together with the O2k-MultiSensor System (there are different requirements for different generations of O2ks). The electrode itself has to be bought from a third party.
Β 
It is sometimes stated that the available H<sub>2</sub>O<sub>2</sub> electrodes may not have the required sensitivity for most ROS studies. WPI states a lower detection limit of "<100 nM" for their 2 mm "macro" electrode and "<10 nM" for their 0.1 mm "mini sensor". The latter value is quite similar to the stated detection limit for fluorescence based methods, however it is not clear if the "mini sensor", probable developed for direct tissue insertion, works well in macroscopic chambers.Β  A direct comparison would be helpful. Please add your experiences!
Β 
Picomolar detection ranges have been reported for a method combining Amplex Red (R) with square wave voltammetry.<ref>Lyon J L, Stevenson K J (2006) Picomolar peroxide detection using a chemically activated redox mediator and square wave voltammetry. Anal. Chem. 78: 8518-8525</ref>
Β 
==References==
<references/>
Β 
Β 
Β 
{{#set:Technical service=NO (amperometric) signal |Technical service=pX signal|Technical service=MultisensorΒ  }}
Β 
__SHOWFACTBOX__
{{Technical service}}

Latest revision as of 19:20, 8 December 2014

Redirect to: