Description | The Titration Set consists of 6x 10 Β΅l, 6x 25 Β΅l, and 2x 50 Β΅l Hamilton glass syringes, and 3 spare plungers for the 10 Β΅l syringes. It is a basic HRR-Accessory included in the O2k-Core Unit. |
---|---|
Product ID | 20820-01 |
Type | O2k, O2k-Module, O2k-Core, Microsyringe, Catalogue |
Link | MiPNet09.12 O2k-Titrations |
Image |
The Titration Set consists of
- 6 Microsyringe\10 mm3 51/0.13 mm 10 mm3
- 6 Microsyringe\25 mm3 51/0.15 mm 25 mm3
- 2 Microsyringe\50 mm3 51/0.15 mm 50 mm3
- 3 Plunger\10 mm3 for 10 mm3 syringe
Handling of the syringes
- Separate inhibitors from substrates.
- Rinse the outside with water immediately after use.
- At the start of a day: rinse the outside with the pure solvent and wash the syringes 3 times with the pure solvent.
- Rinse with pure solvent before use.
- Between two runs during the day: rinse the outside with water.
- At the end of a day: wash 3 x with last used solvent, rinse 3 x with EtOH 100%.
- Storage: ethanol 100% (short), dry (longtime).