O2k signals and output: Difference between revisions

From Bioblast
No edit summary
No edit summary
Line 22: Line 22:




'''NextGen-O2k''': additional channel types are available in the [[NextGen-O2k]]:
'''O2k-Innovation''': additional channel types are available in the [[O2k-Innovation]]:


:::: '''4. Dual-fluo channel'''.
:::: '''4. Dual-fluo channel'''.
Line 79: Line 79:




::: '''Additional output types for the NextGen-O2k'''
::: '''Additional output types for the O2k-Innovation'''


:::: 4. Single-fluo channel - [[NextGen-O2k description|NextGen-O2k]]: Plots of one selected wavelength can be selected for recording. Output type I(1), II(1), III(1) depends on the fluorophore.
:::: 4. Single-fluo channel - [[O2k-Innovation description|O2k-Innovation]]: Plots of one selected wavelength can be selected for recording. Output type I(1), II(1), III(1) depends on the fluorophore.


:::: 5. Dual-fluo channel - [[NextGen-O2k description|NextGen-O2k]]: Plots of two selected wavelengths can be selected for simultaneous recording. Output type I(2), II(2), III(2) depends on the fluorophore.
:::: 5. Dual-fluo channel - [[O2k-Innovation description|O2k-Innovation]]: Plots of two selected wavelengths can be selected for simultaneous recording. Output type I(2), II(2), III(2) depends on the fluorophore.


:::: 6. Spectrophotometric channel - [[NextGen-O2k description|NextGen-O2k]]: Output type IV (three plots for redox states of cytochromes ''aa''<sub>3</sub>, ''b'', and ''c'').
:::: 6. Spectrophotometric channel - [[O2k-Innovation description|O2k-Innovation]]: Output type IV (three plots for redox states of cytochromes ''aa''<sub>3</sub>, ''b'', and ''c'').


:::: '''IV''': Redox states.
:::: '''IV''': Redox states.

Revision as of 12:53, 6 March 2019


high-resolution terminology - matching measurements at high-resolution


O2k signals and output

Description

Three electronic channel types are available in the O2k-MultiSensor system. All channels are available twofold (dual-data), for O2k-Chambers A (left) and B (right), based on numerical signals sent at a fixed data sampling time interval (default: 2 s; range 0.2 s to >10 s).


Reference: O2k-Catalogue: O2k-MultiSensor


MitoPedia O2k and high-resolution respirometry: O2k hardware, DatLab 


O2k signals

up to O2k Series G
O2k Series H
1. O2 channel with electric current (ampere [A]) as the primary signal, which is converted to a voltage (volt [V]) the raw signal. The O2k-FluoRespirometer supports not only the O2 channel (O2 plug), but also the channel types 2 and 3 for the add-on modules.
2a. Fluo channel: The Fluo sensor has electric current (ampere [A]) as the primary signal, which is converted to a voltage (volt [V]) as the raw signal. Therefore, the O2k-Fluo LED2-Module is connected to the Amp plug (up to O2k-Series G) or the Smart Fluo-Sensors are connected to the separate Fluo plug (O2k-Series H upwards).
2b. Amp channel: The amperometric channel of the O2k (Amp plug), has electric current (ampere [A]) as the primary signal, which is converted to a voltage (volt [V]) as the raw signal. The Amp channel is used with the O2k-NO Amp-Module. Up to O2k-Series G, the Amp plug is also used with the O2k-Fluo LED2-Module. Electronically, 2a and 2b are the same channel.
3. pX channel with electric potential (volt [V]) as the primary and raw signal (BNC pX plug for the ion sensitive electrode, ISE; and separate plug for the pX reference electrode, Ref).


O2k-Innovation: additional channel types are available in the O2k-Innovation:

4. Dual-fluo channel.
5. Spectrophotometric channel.


System channels, for instrumental control and monitoring experimental conditions for both O2k-Chamber A and B (single-data). The numberical signals are sent at the same fixed data sampling time interval:
6a. Block temperature: Measured block temperature equivalent to the experimental temperature in the O2k-Chambers at steady state [ยฐC].
6b. Peltier power (system channel): Relative heating or cooling power of the Peltier system.
6c. Barometric pressure (system channel): Measured barometric pressure [kPa].
6d. Ambient temperature (system channel): Measured temperature of room air (O2k-Series G upwards) [ยฐC].
6e. External temperature (system channel): Temperature measured with an independently placed thermistor (O2k-Series G upwards) [ยฐC].


Events carry information on a time point during the experiment, a short event name (text), and a longer string of comments (text):
7. Events: Automatic events (e.g. from the TIP2k) and manually triggered events.


O2k output

The 'output' of a channel is the finally important information, which may be summarized in a data base.


Calibrated signals and derived plots: Three types of output can be derived from the signal of the channels supported by the O2k and displayed as separate plots:
I: Concentration. Examples: O2; Ca2+; concentration is amount of substance per volume [1 ยตM = 1 ยตmolยทdm-3 = 1 nmolยทml-1].
II: Flow, flux, rate. Examples: O2 consumption; H2O2 production; volume-specific flux is the negative or positive time derivative of concentration over time, frequently multiplied by 1,000 [pmolยทs-1ยทml-1]. Volume-specific flux normalized for the cell density [106ยทml-1] is flow [pmolยทs-1ยทml-1]. Volume-specific flux normalized for the mass density of the sample in the O2k-Chamber [mgยทml-1] is mass-specific flux [pmolยทs-1ยทmg-1]. Flux in any respiratory state normalized for flux in a defined reference state is a rate [dimensionless] or flux control ratio.
III: Force, potential. Examples: TMRM and Safranin for mt-membrane potential; redox potential from redox states of cytochromes or NADH .



O2k-Concept
Output types for the O2k-MultiSensor modules
1. O2 channel - all O2k configurations: Output type I and II.
2a. Fluo (Amp) channel - O2k-FluoRespirometer and O2k-Fluo LED2-Module: The focus of the output depends on the specific fluorophore. Output type I, II, or III, or any combination may relate to the primary level of interest.
2b. Amp channel - O2k-NO Amp-Module: Output type I and II.
3a. pX channel - O2k-TPP+ ISE-Module: Output type III.
3b. pX channel - O2k-pH ISE-Module: Output type I and II.


Additional output types for the O2k-Innovation
4. Single-fluo channel - O2k-Innovation: Plots of one selected wavelength can be selected for recording. Output type I(1), II(1), III(1) depends on the fluorophore.
5. Dual-fluo channel - O2k-Innovation: Plots of two selected wavelengths can be selected for simultaneous recording. Output type I(2), II(2), III(2) depends on the fluorophore.
6. Spectrophotometric channel - O2k-Innovation: Output type IV (three plots for redox states of cytochromes aa3, b, and c).
IV: Redox states.


V: Pre-calibrated signals for system channels.


VI: Events.