Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Run DL-Protocol/Set O2 limit"

From Bioblast
Line 1: Line 1:
{{MitoPedia
{{MitoPedia
|abbr=n.a.
|abbr=n.a.
|description='''Protocols''' can be selected in DatLab (DL7.1) which define the sequence of [[Events - DatLab |events]] and [[Marks - DatLab |Marks]]. Instrumental protocols are used for calibrations and instrumental quality control, typically without experimental sample in the incubation medium. [[Substrate-uncoupler-inhibitor titration]] protocols (see [[MitoPedia: SUIT]]) proceed stepwise to activate a sequence of [[coupling control state]]s and [[pathway control state]]s. A specific SUIT protocol can be assigned to O2k-chamber A or B or both. The Titration-Injection-microPump [[TIP2k]] can be programmed to control automatically TIP2k-protocols. Standard protocols are provided. These can be edited and saved under 'Lab protocols' or under the [[User code - DatLab |User]] logged into DatLab.
|description='''DL-Protocols''' can be selected in DatLab (DL7.1) in the pull-down menu 'Protocol': Set DL-Protocol / O2 limit. A DL-Protocol defines the sequence of [[Events - DatLab |events]] and [[Marks - DatLab |Marks]]. Instrumental DL-Protocols are used for calibrations and instrumental quality control, typically without experimental sample in the incubation medium. DL-Protocols for [[Substrate-uncoupler-inhibitor titration]]s (see [[MitoPedia: SUIT]]) proceed stepwise to activate a sequence of [[coupling control state]]s and [[pathway control state]]s. A specific SUIT protocol can be assigned to O2k-chamber A or B or both. The Titration-Injection-microPump [[TIP2k]] can be programmed to control automatically TIP2k-protocols. Standard DL-Protocols are provided. These can be edited and saved under 'Lab protocols' or under the [[User code - DatLab |User]] logged into DatLab.
}}
}}
{{MitoPedia O2k and high-resolution respirometry
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=DatLab
|mitopedia O2k and high-resolution respirometry=DatLab
}}
}}
== Protocol editor ==
:::: The sequence of Events and Marks can be edited with names, definitions, concentrations and titration volumes. After a protocol is edited and saved under a defined name, the protocol has to be selected and assigned to one or both O2k-chambers.
::::: A. Instrumental protocols: calibration and instrumental O2 background test
::::: B. SUIT protocols: linked to titration solutions
::::: C. TIP2k-protocols.
== DL-Protocol ==
== DL-Protocol ==
After selecting a DL-Protocol, the software indicates for the selected O2k-chamber, which [[Events - DatLab |events]] and [[Marks - DatLab |Marks]] have been set (highlighted in green or turquoise for multiple titration event) and which event is to be set next (highlighted in yellow), with the corresponding mark.
After selecting a DL-Protocol, the software indicates for the selected O2k-chamber, which [[Events - DatLab |events]] and [[Marks - DatLab |Marks]] have been set (highlighted in green or turquoise for multiple titration event) and which event is to be set next (highlighted in yellow), with the corresponding mark.
Line 19: Line 13:
== Titrations in SUIT protocols ==
== Titrations in SUIT protocols ==


::::* Stock concentration of substance B in the titration syringe, ''c''<sub>B,in</sub> [mM]
::::* Stock concentration of substance B in the titration syringe, ''c''<sub>B,in</sub> [mM] or [µM] as specified.
::::* Titration volume, ''v'' [µl]
::::* Titration volume, ''v'' [µl]
::::* Chamber volume, ''V'' [µl]
::::* Chamber volume, ''V'' [µl]
::::* Final concentration, ''c''<sub>B</sub> [mM]
::::* Final concentration, ''c''<sub>B</sub> [mM] or [µM] as specified.


:::::: ''c''<sub>B</sub> [mM] = ''c''<sub>B,in</sub> [mM] x ''v'' [µl] / ''V'' [µl]
:::::: ''c''<sub>B</sub> [mM] = ''c''<sub>B,in</sub> [mM] x ''v'' [µl] / ''V'' [µl]
Line 28: Line 22:
::::» [[MiPNet09.12 O2k-Titrations]]
::::» [[MiPNet09.12 O2k-Titrations]]


:::: Example: Malate (M), ''c''<sub>M,in</sub> = 800 mM stock; ''v'' = 5 µl titration volume; ''V'' = 2,000 µl chamber volume
:::: Example: Malate (M), ''c''<sub>M,in</sub> = 800 mM stock; ''v'' = 5 µl titration volume; ''V'' = 2,000 µl chamber volume.


:::::: ''c''<sub>M</sub> [mM] = ''c''<sub>M,in</sub> [mM] x ''v'' [µl] / ''V'' [µl] = 800 mM x 5 µl / 2,000 µl = 2 mM
:::::: ''c''<sub>M</sub> [mM] = ''c''<sub>M,in</sub> [mM] x ''v'' [µl] / ''V'' [µl] = 800 mM x 5 µl / 2,000 µl = 2 mM

Revision as of 08:58, 30 June 2017


high-resolution terminology - matching measurements at high-resolution


Run DL-Protocol/Set O2 limit

Description

DL-Protocols can be selected in DatLab (DL7.1) in the pull-down menu 'Protocol': Set DL-Protocol / O2 limit. A DL-Protocol defines the sequence of events and Marks. Instrumental DL-Protocols are used for calibrations and instrumental quality control, typically without experimental sample in the incubation medium. DL-Protocols for Substrate-uncoupler-inhibitor titrations (see MitoPedia: SUIT) proceed stepwise to activate a sequence of coupling control states and pathway control states. A specific SUIT protocol can be assigned to O2k-chamber A or B or both. The Titration-Injection-microPump TIP2k can be programmed to control automatically TIP2k-protocols. Standard DL-Protocols are provided. These can be edited and saved under 'Lab protocols' or under the User logged into DatLab.

Abbreviation: n.a.


MitoPedia O2k and high-resolution respirometry: DatLab 

DL-Protocol

After selecting a DL-Protocol, the software indicates for the selected O2k-chamber, which events and Marks have been set (highlighted in green or turquoise for multiple titration event) and which event is to be set next (highlighted in yellow), with the corresponding mark.

DL-Protocol will inform you if a titration event is missed or a repetition of a non multiple titration ocurred (highlighted in red).

Titrations in SUIT protocols

  • Stock concentration of substance B in the titration syringe, cB,in [mM] or [µM] as specified.
  • Titration volume, v [µl]
  • Chamber volume, V [µl]
  • Final concentration, cB [mM] or [µM] as specified.
cB [mM] = cB,in [mM] x v [µl] / V [µl]
» MiPNet09.12 O2k-Titrations
Example: Malate (M), cM,in = 800 mM stock; v = 5 µl titration volume; V = 2,000 µl chamber volume.
cM [mM] = cM,in [mM] x v [µl] / V [µl] = 800 mM x 5 µl / 2,000 µl = 2 mM