Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Chamber assembly"

From Bioblast
(created page)
Β 
Tag: Redirect target changed
Β 
(10 intermediate revisions by 5 users not shown)
Line 1: Line 1:
The assembly of the chamber and a procedure to position the glass chamber very accurately is described in [http://www.oroboros.at/index.php?id=o2k-start| MiPNet12.06].
#REDIRECT [[O2k-chamber#Chamber_assembly]]
A wrongly positioned glass chamber can lead to a huge variety of problems, some obvious but others very difficult to understand.
==Possible Problems==
* chamber leaking liquids (medium is leaking out of the chamber)
* oxygen leak of the chamber (strongly negative intercept in the [[instrumental background test]]
* unstable O2 signal
* unexpected high or low O2 signal
Β 
==Trouble shooting procedure==
If the problem is ob obviously related to the chamber proceed to solutions.|
Otherwise followΒ  [[Locating a problem]]
If the problem is not located on the [[POS]] or the [[POS connector]] try solving the problem by reassembling the chamber before concluding that the [[main unit]] is defect.
Β 
==Solutions==
* Disassemble the chamber by reversing the instructions in [http://www.oroboros.at/index.php?id=o2k-start| MiPNet12.06].
* Check for visible damages on the [[glass chamber]], on the [[POS seal tip]] and all o-rings (on chamber holders or on stoppers). If a damage is visible replace the part or consult OROBOROS Instruments. If there is a deposit visible in the glass chamber see [[Cleaning the glass chamber]].
* Reassemble the chamber following exactly the description in [http://www.oroboros.at/index.php?id=o2k-start| MiPNet12.06].

Latest revision as of 12:04, 18 October 2020