| {{Technical service}}Β
| |
| '''Problem:''' After removing [[OroboPOS-Connector]], [[OroboPOS-Holder]], and [[O2k-Chamber Holder]]
| |
| from the O2k , the [[O2k-Chamber]]Β cannot be removed. It remains stuck even after washing with water.
| |
|
| |
| '''Solution''': The reason for this problem is very likely medium that has been spilled and acts now as a glue between glass chamber and copper block.Β
| |
| Wash with plenty of warm water and try to remove the chamber while the O2k is switched on and heated up to 37 Β°C. You can try GENTLY to use the sensor holder (not the sensor connector) to try to push up the glass chamber (like you do during [[chamber assembly]]) - while the O2k is warm and with plenty of warm water.Β Since the sensors are removed you also can use e.g. 40Β° (or a little bit more) for the washing / removing attempts.
| |
|
| |