The two chamber approach of the O2k has among other benefits also advantages for [[troubleshooting]]. By switching components between sides it usually is possible to locate a problem specifically at one component. This component can then be serviced by the user or replaced by OROBOROS Instruments without the need to ship the entire oxygraph.
#REDIRECT [[O2k-Open Support]]
 
 
== Locating a Problem by Switching Components ==
 
It is always important to access the performance of the instrument in a well defined manner. Therefore, at each stage of this process a full [[Sensor Test]] has to be done. In such a sensor test all hardware problems are visible. Biological experiments are not suitable for trouble shooting purposes. For troubleshooting it is very helpful to observe the [[Raw Signal]] rather than a calibrated signal.
 
1.) Perform a [[Sensor Test]]
 
2.) Switch the [[sensor heads]] ([[POS]]) between right and left side. This is also a good opportunity to clean and apply contact oil to the gold pin and thread connecting the [[sensor connectors]] and proper [[sensor heads]]. When disconnecting a sensor from the sensor connector remember to observe the guidelines to prevent damage by ESD, see [http://www.oroboros.at/index.php?id=esd-protection [MiPNet14.1]]
 
3.) Repeat the full [[Sensor Test]]. If the problem moved together with the sensor from one side to the other, the problem is located in the sensor head (POS)
 
4.) If the problem remained on the same side, switch the [sensor connectors] together with the attached sensors between right and left side.
 
5.) Repeat the full [[Sensor Test]]. If the problem moved together with the sensor connector, it is located at the sensor connector. If the problem still stays on the same side check carefully the Datlab settings for this side. There may be wrong settings for [[gain]], polarization voltage or scaling. Or there may be a wrong calibration for this channel (this is easily detected by observing the [[Raw Signal]]). When these problems have been ruled out an electronic defect in the main unit for this channel must be considered.
 
{{#set:topic=sensor|topic=sensor connector|article type=instruction}}