Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "MiPNet19.18D O2k-Series G and DatLab 6: Calibration"

From Bioblast
Line 23: Line 23:
}}
}}
{{Labeling
{{Labeling
|instruments=Oxygraph-2k, MultiSensor pH; TPP; NO, DatLab Software; Separate Application, Method, Protocol; Manual
|instruments=Oxygraph-2k, Protocol, MiPNet-Publication
|articletype=Protocol; Manual, MiPNet-online Publication
|articletype=Protocol; Manual, MiPNet-online Publication
}}
}}
Line 29: Line 29:


MiPNet08.07, MiPNet09.08 and MiPNet02.01 (DatLab 2) is integrated into MiPNet12.08.
MiPNet08.07, MiPNet09.08 and MiPNet02.01 (DatLab 2) is integrated into MiPNet12.08.
[[Category:OroboroPedia]]

Revision as of 17:37, 13 May 2011

Publications in the MiPMap
Gnaiger E. Oxygen and pX Calibration DatLab4.3. Mitochondr. Physiol. Network 12.08.

Β» MiPNet12.08

Gnaiger E (

) Mitochondr. Physiol. Network

Abstract: Accurate calibration of the oxygen sensor depends on (1) calibration solutions prepared at known oxygen partial pressures, as achieved in the Oxygraph-2k (O2k) at defined temperature, continuously recorded total gas pressure (barometric pressure), and thermodynamic equilibrium between the gas and aqueous phase; (2) high stability of the signal of the polarographic oxygen sensor (POS), tested for sufficiently long periods of time; (3) linearity of signal output with oxygen pressure, achieved with the POS in the range between oxygen saturation and zero oxygen pressure; and (4) application of accurate oxygen solubilities for aqueous solutions for the conversion of partial oxygen pressure into oxygen concentration MiPNet06.03. The standard oxygen calibration procedure is described for O2k high-resolution respirometry with the automatic calibration routine of DatLab 4. β€’ Keywords: HRR-DatLab

β€’ O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels:





HRR: Oxygraph-2k, Protocol"Protocol" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property., MiPNet-Publication"MiPNet-Publication" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property. 


Info

MiPNet08.07, MiPNet09.08 and MiPNet02.01 (DatLab 2) is integrated into MiPNet12.08.