Line 17: Line 17:
oxygen pressure; and (4) application of accurate [[oxygen solubility]] for
oxygen pressure; and (4) application of accurate [[oxygen solubility]] for
aqueous solutions for the conversion of partial [[oxygen pressure]] into
aqueous solutions for the conversion of partial [[oxygen pressure]] into
oxygen concentration ([[MiPNet06.03]]). The standard oxygen calibration
[[oxygen concentration]] ([[MiPNet06.03]]). The standard oxygen calibration
procedure is described for O2k high-resolution respirometry with the
procedure is described for O2k high-resolution respirometry with the
automatic calibration routine of DatLab.
automatic calibration routine of DatLab.
Line 30: Line 30:
|instruments=Oxygraph-2k, O2k-Manual, Protocol
|instruments=Oxygraph-2k, O2k-Manual, Protocol
|additional=O2k-SOP, DatLab
|additional=O2k-SOP, DatLab
|articletype=Protocol; Manual, MiPNet-online Publication
}}
}}
== Further information ==
== Further information ==
:>> [http://www.oroboros.at/index.php?O2k-o2calibration  MiPNet12.08]
:>> [http://www.oroboros.at/index.php?O2k-o2calibration  MiPNet12.08]

Revision as of 10:01, 17 October 2013

Publications in the MiPMap
Gnaiger E (2013) Oxygen and pX (pH) Calibration by DatLab. Mitochondr Physiol Network 12.08.

» Open Access-MiPNet12.08.pdf, Versions

O2k-Manual contents

OROBOROS (2013) Mitochondr Physiol Network

Abstract: O2k-Manual Chapter D

Accurate calibration of the oxygen sensor depends on (1) calibration solutions prepared at known oxygen partial pressures, as achieved in the Oxygraph-2k (O2k) at defined temperature, continuously recorded total gas pressure (barometric pressure), and thermodynamic equilibrium between the gas and aqueous phase; (2) high stability of the signal of the polarographic oxygen sensor (POS), tested for sufficiently long periods of time; (3) linearity of signal output with oxygen pressure, achieved with the POS in the range between oxygen saturation and zero oxygen pressure; and (4) application of accurate oxygen solubility for aqueous solutions for the conversion of partial oxygen pressure into oxygen concentration (MiPNet06.03). The standard oxygen calibration procedure is described for O2k high-resolution respirometry with the automatic calibration routine of DatLab.

>> O2k-Manual contents
>> OROBOROS Oxygraph-2k

Keywords: HRR, DatLab

O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual, Protocol"Protocol" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property. 

O2k-SOP, DatLab 


Further information

>> MiPNet12.08
>> MiPNet08.07, MiPNet09.08 and MiPNet02.01 (DatLab 2) are integrated into MiPNet12.08.