Difference between revisions of "MiPNet19.18D O2k-Series G and DatLab 6: Calibration"

From Bioblast
Line 1: Line 1:
{{Publication
{{Publication
|title=Gnaiger E (2013) Oxygen and pX (pH) Calibration by DatLab. Mitochondr Physiol Network 12.08. Version 07.
|title=Gnaiger E (2014) Oxygen calibration by DatLab. Mitochondr Physiol Network 12.08(08): 1-10.
|info=[[Media:MiPNet12.08 O2k-Calibration.pdf|'''Open Access-MiPNet12.08.pdf''']], [http://www.bioblast.at/index.php/File:MiPNet12.08_O2k-Calibration.pdf Versions] [[Image:O2k-Manual.jpg|right|150px|link=http://www.oroboros.at/?o2k-manual|O2k-Manual contents]]
|info=[[Media:MiPNet12.08 O2k-Calibration.pdf|'''Open Access-MiPNet12.08.pdf''']], [http://www.bioblast.at/index.php/File:MiPNet12.08_O2k-Calibration.pdf Versions] [[Image:O2k-Manual.jpg|right|150px|link=http://www.oroboros.at/?o2k-manual|O2k-Manual contents]]
|authors=OROBOROS
|authors=OROBOROS
|year=2013
|year=2014
|journal=Mitochondr Physiol Network
|journal=Mitochondr Physiol Network
|abstract='''O2k-Manual Chapter D'''  
|abstract='''O2k-Manual Chapter D'''  
Line 35: Line 35:
:>> [http://www.oroboros.at/?O2k-O2calibration  Further detail? Consult OROBOROS]
:>> [http://www.oroboros.at/?O2k-O2calibration  Further detail? Consult OROBOROS]


* Version 06: 2012-10-24
* Version 01: 2007
* MiPNet08.07, MiPNet09.08 and MiPNet02.01 (DatLab 2) are integrated into MiPNet12.08.





Revision as of 18:50, 16 February 2014

Publications in the MiPMap
Gnaiger E (2014) Oxygen calibration by DatLab. Mitochondr Physiol Network 12.08(08): 1-10.

» Open Access-MiPNet12.08.pdf, Versions

O2k-Manual contents

OROBOROS (2014) Mitochondr Physiol Network

Abstract: O2k-Manual Chapter D

Accurate calibration of the oxygen sensor depends on (1) calibration solutions prepared at known oxygen partial pressures, as achieved in the Oxygraph-2k (O2k) at defined temperature, continuously recorded total gas pressure (barometric pressure), and thermodynamic equilibrium between the gas and aqueous phase; (2) high stability of the signal of the polarographic oxygen sensor (POS), tested for sufficiently long periods of time; (3) linearity of signal output with oxygen pressure, achieved with the POS in the range between oxygen saturation and zero oxygen pressure; and (4) application of accurate oxygen solubility for aqueous solutions for the conversion of partial oxygen pressure into oxygen concentration (MiPNet06.03). The standard oxygen calibration procedure is described for O2k high-resolution respirometry with the automatic calibration routine of DatLab.

>> O2k-Manual: Overall contents
>> Product: OROBOROS Oxygraph-2k, O2k-Catalogue

Keywords: HRR, DatLab

O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual, Protocol"Protocol" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property. 

O2k-SOP, DatLab 

Further information

>> Further detail? Consult OROBOROS