Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "MiPNet21.07 MitoFit DatLab PT"

From Bioblast
Line 38: Line 38:


# DatLab proficiency test instructions: [[Media:DatLabanalysisTest_instructions.pdf|Instructions]]
# DatLab proficiency test instructions: [[Media:DatLabanalysisTest_instructions.pdf|Instructions]]
# Experimental data file: [[Media:file_exp1_xx.DLD|file_exp1_xx.DLD]]
# Experimental data file: [[Media:file_exp1_xx.DLD|2015-09-24 P1-01.DLD]]
# Template: [[Media:Template1 xx.xls|Template1 xx.xls]]
# Template: [[Media:Template1 xx.xls|NS 1PM 2D 3G 4c 5S 6U 7Rot 8Ama.xls]]
# Calibration and Background file: [[Media:file_bgr1_xx.DLD|file_bgr1_xx.DLD]]
# Calibration and Background file: [[Media:file_bgr1_xx.DLD|2015-03-30 P1-01.DLD]]

Revision as of 17:35, 16 February 2016

Publications in the MiPMap
O2k-Protocols
MitoFit DatLab proficiency test. - »Bioblast pdf«

» Versions

OROBOROS (2016-01-21) Mitochondr Physiol Network

Abstract: Krumschnabel G, Plattner C, Gnaiger (2016) MitoFit DatLab proficiency test. Mitochondr Physiol Network 21.07(01):1-##.

The MitoFit DatLab analysis test can be considered as a first-level 'dry' MitoFit Proficiency Test for evaluation of inter-individual and inter-laboratory reproducibility and accuracy of DatLab analysis of the same data files. Proficiency will be tested for (1) setting Marks for analysis of flux; (2) O2 calibration at air saturation and zero O2 concentration; (3) analysis of instrumental O2 background.



» DatLab


O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Instruments;methods 


Organism: Human 

Preparation: Permeabilized cells 


Coupling state: LEAK, OXPHOS, ETS"ETS" is not in the list (LEAK, ROUTINE, OXPHOS, ET) of allowed values for the "Coupling states" property. 

HRR: Oxygraph-2k, O2k-Protocol 

MitoFitPublication 

Aims

  1. Obtain a measure of noise due to variability in setting marks on 'flux' when evaluating flux in different respiratory states.
  2. Guide the user to the DatLab-Excel template for analysis.
  3. Proficiency test: Two-point O2 calibration, with O2 solubility factor.
  4. Obtain a measure of noise of instrumental O2 background parameters due to variability in setting marks on 'flux'.
  5. Proficiency test: Application of instrumental O2 background to analysis of flux.


Files to download:

  1. DatLab proficiency test instructions: Instructions
  2. Experimental data file: 2015-09-24 P1-01.DLD
  3. Template: NS 1PM 2D 3G 4c 5S 6U 7Rot 8Ama.xls
  4. Calibration and Background file: 2015-03-30 P1-01.DLD