The MitoFit DatLab analysis test can be considered as a first-level 'dry' MitoFit Proficiency Test for evaluation of inter-individual and inter-laboratory reproducibility and accuracy of DatLab analysis of the same data files. Proficiency will be tested for (1) setting Marks for analysis of flux; (2) O2 calibration at air saturation and zero O2 concentration; (3) analysis of instrumental O2 background and copying of parameters for flux analysis. Β 
The MitoFit DatLab analysis test can be considered as a first-level 'dry' MitoFit Proficiency Test for evaluation of inter-individual and inter-laboratory reproducibility and accuracy of DatLab analysis of the same data files. Proficiency will be tested for (1) setting Marks for analysis of flux; (2) O2 calibration at air saturation and zero O2 concentration; (3) analysis of instrumental O2 background and copying of parameters for flux analysis. Β 


===Aims===
===Aims===
Line 41: Line 42:
===Files to download:===
===Files to download:===


# Instructions: [[File:DatLabanalysisTest_instructions.pdf]]
# DatLab proficiency test instructions: [[Media:DatLabanalysisTest_instructions.pdf|Instructions]]
# Experimental data file: [[File:file_exp1_xx.DLD]]
# Experimental data file: [[Media:file_exp1_xx.DLD|file_exp1_xx.DLD]]
# Template: [[File:Template1 xx.xls]]
# Template: [[Media:Template1 xx.xls|Template1 xx.xls]]
# Calibration and Background file: [[File:file_bgr1_xx.DLD]]
# Calibration and Background file: [[Media:file_bgr1_xx.DLD|file_bgr1_xx.DLD]]

Revision as of 11:41, 28 January 2016

Publications in the MiPMap
O2k-Protocols
MitoFit DatLab proficiency test. - Β»Bioblast pdfΒ«

Β» Versions

OROBOROS (2016-01-21) Mitochondr Physiol Network

Abstract: Krumschnabel G, Plattner C, Gnaiger (2016) MitoFit DatLab proficiency test. Mitochondr Physiol Network 21.07(01):1-##.

Β» Instrument: O2k, DatLab, O2k-Catalogue
  • Page in prep.


β€’ O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Instruments;methods 


Organism: Human 

Preparation: Permeabilized cells 


Coupling state: LEAK, OXPHOS, ETS"ETS" is not in the list (LEAK, ROUTINE, OXPHOS, ET) of allowed values for the "Coupling states" property. 

HRR: Oxygraph-2k, O2k-Protocol 

O2k-SOP, MitoFitPublication 


Abstract:

The MitoFit DatLab analysis test can be considered as a first-level 'dry' MitoFit Proficiency Test for evaluation of inter-individual and inter-laboratory reproducibility and accuracy of DatLab analysis of the same data files. Proficiency will be tested for (1) setting Marks for analysis of flux; (2) O2 calibration at air saturation and zero O2 concentration; (3) analysis of instrumental O2 background and copying of parameters for flux analysis.


Aims

  1. Obtain a measure of noise due to variability in setting marks on 'flux' when evaluating flux in different respiratory states.
  2. Guide the user to the DatLab-Excel template for analysis.
  3. Proficiency test: Two-point O2 calibration, with O2 solubility factor.
  4. Obtain a measure of noise of instrumental O2 background parameters due to variability in setting marks on 'flux'.
  5. Proficiency test: Application of instrumental O2 background to analysis of flux.


Files to download:

  1. DatLab proficiency test instructions: Instructions
  2. Experimental data file: file_exp1_xx.DLD
  3. Template: Template1 xx.xls
  4. Calibration and Background file: file_bgr1_xx.DLD