O2k-Barometric Pressure Transducer: Difference between revisions
From Bioblast
No edit summary |
No edit summary |
||
Line 5: | Line 5: | ||
|info=[[O2k-Core]] | |info=[[O2k-Core]] | ||
}} | }} | ||
:::: SI unit: | :::: SI unit: Pa = Jยทm<sup>-3</sup> | ||
:::: 1 atm = 760 mmHg = 101.325 kPa | :::: 1 atm = 760 mmHg = 101.325 kPa | ||
Revision as of 09:16, 12 March 2021
O2k-Barometric Pressure Transducer
Description | O2k-Barometric Pressure Transducer, A/D converter and digital output to DatLab for continuous recording of barometric pressure [kPa or mmHg], integrated into the air calibration of the OroboPOS (MiPNet06.03 POS-calibration-SOP). Integral component of the O2k-Main Unit. The warranty on the accuracy of the signal obtained from the O2k-Barometric Pressure Transducer expires within three years. |
---|---|
Product ID | 20050 |
Type | O2k-Main Unit |
Link | O2k-Core |
Image |
- SI unit: Pa = Jยทm-3
- 1 atm = 760 mmHg = 101.325 kPa
Accuracy
- Quick accuracy check for the barometric pressure transducer (Mario Faschung, 2013-10-21):
- 4 O2k: average pressure reading = 95.5 kPa, maximal reading 95.5 kPa, minimal reading 95.4 kPa
- Nearest metrological station (Innsbruck airport, altitude: 585 m above sea level): pressure = 95.5 kPa