== O2k-OroboPOS-Holder: Leaky chamber due to damaged OroboPOS-Holder== | == O2k-OroboPOS-Holder: Leaky chamber due to damaged OroboPOS-Holder== | ||
:::* In order not to experience a leakage after chamber reassembly, make sure that all the parts are in good conditions, especially the OroboPOS-Holder ย | :::* In order not to experience a leakage after chamber reassembly, make sure that all the parts are in good conditions, especially the OroboPOS-Holder . | ||
:::* The OroboPOS-Holder may experience some damage due to the not proper alignment with the thread ย | :::* The OroboPOS-Holder may experience some damage due to the not proper alignment with the thread. | ||
:::* The OroboPOS-Holder may experience some damage due to the not proper alignment with the chamber | :::* The OroboPOS-Holder may experience some damage due to the not proper alignment with the chamber. | ||
:::* The OroboPOS-Holder may experience some damage due to frequent and bad performed mounting and dismounting procedures | :::* The OroboPOS-Holder may experience some damage due to frequent and bad performed mounting and dismounting procedures. | ||
:::* In case your OroboPOS-Holder already presents dents on the surfaces ( see pictures below) you may experience leakage | :::* In case your OroboPOS-Holder already presents dents on the surfaces ( see pictures below) you may experience leakage. Although sanding the edge of the OroboPOS-Holder with sand paper may solve the problem, replacement of a damaged OroboPOS-Holder is recommended. | ||
[[File:OroboPOS-Holder002.jpg|left|200px]] | [[File:OroboPOS-Holder002.jpg|left|200px]] | ||
Line 28: | Line 23: | ||
[[File:OroboPOS-Holder004.jpg|centre|200px]] | [[File:OroboPOS-Holder004.jpg|centre|200px]] | ||
* Left picture: brand new OroboPOS-Holder, no damage on the thred, no damage on the edge | * Left picture: brand new OroboPOS-Holder, no damage on the thred, no damage on the edge. | ||
* Center picture: used OroboPOS-Holder, almost no damage on the thred, the edge is slightly damaged ย | * Center picture: used OroboPOS-Holder, almost no damage on the thred, the edge is slightly damaged. | ||
* Right picture: heavly used OroboPOS-Holder, the thred is slightly damaged, serious damage on the edge of the aperture | * Right picture: heavly used OroboPOS-Holder, the thred is slightly damaged, serious damage on the edge of the aperture. | ||
:::* | :::* ''More details are shown in this video'' - ยป[https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]ยซ | ||
{{MitoPedia O2k and high-resolution respirometry | {{MitoPedia O2k and high-resolution respirometry | ||
|mitopedia O2k and high-resolution respirometry=O2k hardware | |mitopedia O2k and high-resolution respirometry=O2k hardware | ||
}} | }} |
Revision as of 19:29, 4 March 2020
- high-resolution terminology - matching measurements at high-resolution
O2k-OroboPOS-Holder MitoPedia
Description
The OroboPOS-Holder (POS-Holder), made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding the OroboPOS-Connector with the OroboPOS sensor head (POS) to the O2k-Chamber, and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-Chamber with the OroboPOS-Seal Tip. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.
Two units of this item are standard components mounted on the O2k-Main Unit.
Reference: OroboPOS-Holder
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
O2k-OroboPOS-Holder: Leaky chamber due to damaged OroboPOS-Holder
- In order not to experience a leakage after chamber reassembly, make sure that all the parts are in good conditions, especially the OroboPOS-Holder .
- The OroboPOS-Holder may experience some damage due to the not proper alignment with the thread.
- The OroboPOS-Holder may experience some damage due to the not proper alignment with the chamber.
- The OroboPOS-Holder may experience some damage due to frequent and bad performed mounting and dismounting procedures.
- In case your OroboPOS-Holder already presents dents on the surfaces ( see pictures below) you may experience leakage. Although sanding the edge of the OroboPOS-Holder with sand paper may solve the problem, replacement of a damaged OroboPOS-Holder is recommended.
- Left picture: brand new OroboPOS-Holder, no damage on the thred, no damage on the edge.
- Center picture: used OroboPOS-Holder, almost no damage on the thred, the edge is slightly damaged.
- Right picture: heavly used OroboPOS-Holder, the thred is slightly damaged, serious damage on the edge of the aperture.
- More details are shown in this video - ยปChamber service Insert O2k Chamberยซ