== Static POS calibration ==
#REDIRECT [[MiPNet06.03 POS-calibration-SOP]]
ย 
'''Static POS calibration''' = Two-point calibration of the [[polarographic oxygen sensor]].
ย 
DatLab converts the [[raw signal]] to oxygen concentrationsย  by applying conversion factors determined during a POS calibration comprising an [[air calibration]] and a [[zero calibration]]. For each [[Gain (O2 channel) | gain]] a different set of calibration values is necessary. Frequently, assumed hardware problems turn out to be a simple calibration problem. This can easily be detected by observing the [[Raw signal]].
Please see the Calibration Manual[[http://www.oroboros.at/index.php?id=o2k-o2calibration MiPNet12.08]]
ย 
ย 
== Dynamic POS calibration ==
ย 
'''Dynamic POS calibration''' = Calibration of the [[response time of the POS]].
ย 
ย 
ย 
ย 
{{#set:Technical service=POS calibration|Technical service=Oxygen solubility}}
__SHOWFACTBOX__
{{troubleshooting}}