|
ย |
(5 intermediate revisions by 3 users not shown) |
Line 1: |
Line 1: |
| {{Technical support}}
| | #REDIRECT [[MiPNet06.03 POS-calibration-SOP]] |
| ย | |
| ย | |
| == Static POS calibration ==
| |
| ย | |
| '''Static POS calibration''' = Two-point calibration of the [[polarographic oxygen sensor]].
| |
| ย | |
| DatLab converts the [[raw signal]] to oxygen concentrationsย by applying conversion factors determined during a POS calibration comprising an [[air calibration]] and a [[zero calibration]]. For each [[Gain (O2 channel) | gain]] a different set of calibration values is necessary. Frequently, assumed hardware problems turn out to be a simple calibration problem. This can easily be detected by observing the [[Raw signal]].
| |
| Please see the Calibration Manual [[MiPNet19.01D O2k-Calibration]].
| |
| ย | |
| ย | |
| == Dynamic POS calibration ==
| |
| ย | |
| '''Dynamic POS calibration''' = Calibration of the [[response time of the POS]].
| |
| ย | |
|
| |
| ย | |
| ย | |
| ย | |
| {{#set:Technical service=POS calibration}}
| |