== Static POS calibration ==
ย 
'''Static POS calibration''' = Two-point calibration of the [[polarographic oxygen sensor]].
ย 
DatLab converts the [[raw signal]] to oxygen concentrationsย  by applying conversion factors determined during a POS calibration comprising an [[air calibration]] and a [[zero calibration]]. For each [[Gain (O2 channel) | gain]] a different set of calibration values is necessary. Frequently, assumed hardware problems turn out to be a simple calibration problem. This can easily be detected by observing the [[Raw signal]].
Please see the Calibration Manual [[MiPNet19.01D O2k-Calibration]].
ย 
ย 
== Dynamic POS calibration ==
ย 
'''Dynamic POS calibration''' = Calibration of the [[response time of the POS]].
ย 