| == Static POS calibration ==
| | {{Person}} |
| ย | |
| '''Static POS calibration''' = Two-point calibration of the [[polarographic oxygen sensor]].
| |
| ย | |
| DatLab converts the [[raw signal]] to oxygen concentrationsย by applying conversion factors determined during a POS calibration comprising an [[air calibration]] and a [[zero calibration]]. For each [[Gain (O2 channel) | gain]] a different set of calibration values is necessary. Frequently, assumed hardware problems turn out to be a simple calibration problem. This can easily be detected by observing the [[Raw signal]].
| |
| Please see [[http://www.oroboros.at/index.php?id=o2k-o2calibration MiPNet12.08]]
| |