POS connector (technical service)

From Bioblast
Revision as of 09:11, 29 September 2010 by Fasching Mario (talk | contribs) (link to Cleaning the POS conector)

For general and purchasing information see POS connector.

The two electronic connections of the POS connector (to the POS and to the Main unit) make a POS connector that has been disconnected from the POS or from the Main unit or from both particular sensitive to damage by ESD. Therefore it is of special importance to follow the guidelines for ESD protection [MiPNet14.1] whenever handling a POS connector.

Functions

The POS connector has several different functions:

  • It mechanically presses the POS against the glass chamber, ensuring a tight fit.
  • It connects the POS electronically to the main unit transmitting the signal.
  • In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the Main unit. For this reason POS connectors for Series A-C and for Series D are NOT interchangeable.

Possible Problems

Nearly all problems usually associated with the POS may in fact be caused by the POS connectors, e.g.

  • unstable O2 signal
  • unstable O2 flux
  • high O2 signal at air saturation
  • high O2 signal at zero oxygen


Performing the Sensor Test

Confirm the problem by doing a Sensor test in the absence of biological material

Option 1.A: No problem was visible in the Sensor Test

check the following DatLab settings

If the problem is not visible when observing the Raw signal but is visible when observing the calibrated oxygen signal, then there is probable a problem with the POS calibration.

Option 1.B: The problem was visible in the Sensor Test

  • proceed to Localizing the Problem

Localizing the Problem

Locate the Problem

Option 2.A: The problem was localized on the POS connector

Solutions 2.A Follow the instructions for cleaning and applying contact oil as described in the sensor service manual [MiPNet08.04] and in more detail at Cleaning the POS connector. If you have also done a Sensor service or changed the POS membranes allow the oxygraph to run over night before repeating the Sensor test. If all service precautions fail to solve the problem, or a new POS connector has to be applied.

Option 2.B: The problem was not localized on the POS connector

Solutions 2.B If the problem was localized on an other component follow the instructions for this component. If you could not locate the problem contact OROBOROS Instruments.




Template:Technical service