The response time of the [[Polarographic oxygen sensor]] is assessed in the [[Stirrer test]] that is part of a full [[Sensor test]]. A quantitative interpretation of the stirrer test will yield the time constant of the sensor. Such a quantitative calculation is indeed necessary for kinetic work. However, to access the suitability of a sensor for standard steady state work each user will soon gain the necessary experience to judge the outcome of a stirrer test. Beginners can have a look at the example [[sensor test]] available for download [http://www.oroboros.at/index.php?id=sensor_test here]. | The response time of the [[Polarographic oxygen sensor]] is assessed in the [[Stirrer test]] that is part of a full [[Sensor test]]. A quantitative interpretation of the stirrer test will yield the time constant of the sensor. Such a quantitative calculation is indeed necessary for kinetic work. However, to access the suitability of a sensor for standard steady state work each user will soon gain the necessary experience to judge the outcome of a stirrer test. Beginners can have a look at the example [[sensor test]] available for download [http://www.oroboros.at/index.php?id=sensor_test here]. | ||
Line 7: | Line 9: | ||
{{#set:Technical service=O2 sensor|Technical service=O2 signal}} | {{#set:Technical service=O2 sensor|Technical service=O2 signal}} | ||
[[Category:Technical service]] | |||
__SHOWFACTBOX__ | __SHOWFACTBOX__ | ||
Revision as of 16:58, 13 February 2015
The response time of the Polarographic oxygen sensor is assessed in the Stirrer test that is part of a full Sensor test. A quantitative interpretation of the stirrer test will yield the time constant of the sensor. Such a quantitative calculation is indeed necessary for kinetic work. However, to access the suitability of a sensor for standard steady state work each user will soon gain the necessary experience to judge the outcome of a stirrer test. Beginners can have a look at the example sensor test available for download here.
For how to improve the response time of a POS, see Slow responding sensor.
The response time of the sensor is just one of several factor determining the time resolution of an experiment.