Line 143: Line 143:


As  expected some traditional KCl based media, similar to those used in  ... showed a far higher sensitivity against h2O2 addition then MiR05,  while some experiments indicated a lower respiration.
As  expected some traditional KCl based media, similar to those used in  ... showed a far higher sensitivity against h2O2 addition then MiR05,  while some experiments indicated a lower respiration.
=== Media 5: Comparison MiR05Cr to MiRK03Cr ===
'''Conditions:'''
''Instrumental settings:''
*Amp.Gain 1000
*Light intesity variable 1mA, 2mA and 5mA
*Configuration O2K -Sensor always the same: Oxygraph AB: Sensors B-0111/B-0082 Oxygraph CD: Sensors B-0119/B-0117 Oxygraph EF: Sensors A-0004/B-0112 Oxygraph GH: Sensors A-0087/A-0088
''Media:''
MiR05Cr and 
MiRK03Cr:
KCl 130 mM,
HEPES free acid 20 mM,
KH2PO4 10 mM, 
MgCl2 3 mM,
EGTA 0.5 mM, 
BSA : 0.1%
pH=7
add 120 mg Creatine to 40 ml buffer
'''Titration:'''
10µl Amplex UltraRed (conc:1mM/ml -> 5µM/ml in chamber)
4µl HRP (conc:500U/ml -> 1U/ml in chamber) 
2*10 µl of 56,32 µM H2O2 solution
'''Results/Figures:'''
Drift and Sensitivity at least as good as MitoOx2 (=MiRK02) see Fig. 6 and 7 above




Line 149: Line 178:
[[File:SD drift at diff. H2O2 conc.png]]
[[File:SD drift at diff. H2O2 conc.png]]


no dependence on light intensity ( 1 to 5 mA) 
[[File:Light int. vs. calib.drift in MiR05Cr.png]]
[[File:Light int. vs. calib.drift in MiR05Cr.png]]


[[File:Light int. vs.calib. drift in MiRK03Cr.png]]
[[File:Light int. vs.calib. drift in MiRK03Cr.png]]
Fig.14 and 15
Files: can be found in P:\3D_Mario\1_development_Fluorescence\amplex\MiRK03\DatLab data
MiRK03Cr:
1 mA: 2013-04-30 AB02/CD02/EF02/GH02
2 mA: 2013-06-18 AB01/CD01/EF01/GH01
5 mA: 2013-06-18 AB02/CD02/EF02/GH02
MiR05Cr:
1mA: 2013-04-30 AB01/CD01/EF01/GH01
2mA: 2013-06-18 AB03/CD03/EF03/GH03
5mA: 2013-07-02 AB01/CD01/ 2013-06-28 EF01/GH01
Evaluation of  respirometry in MiR05Cr versus MiRK03Cr (N=2):
 
Fig.16 and 17
Files to be found in P:\3C_Projects_Data\2013_all data\2013_Mouse Heart\Amplex red
2013-01-22 EF02.DLD (MiR05Cr) and 2013-01-22 GH02.DLD (MiRK03Cr)
2013-01-29 EF02.DLD (MiR05Cr) and 2013-01-29 GH02.DLD (MiRK03Cr)
2013-02-05 EF02.DLD (MiR05Cr) and 2013-02-05 GH02.DLD (MiRK03Cr)_ not included in Figure 
To do:? with swiched chambers with and without Creatine in low light intensity
Fontanam 16:22, 6 August 2013 (CEST)


[[File:Drift signal ratios.png]]
[[File:Drift signal ratios.png]]

Revision as of 09:02, 18 September 2013

Different Brands

Trade Mark Manufacturer/ Distributor
price [€/mg]
product id
Amplex Red
Invitrogen
37.4
A12222
Amplex Ultra Red
Invitrogen
48.8
A36006
Ampliflu Red
Sigma
18.5
90101
Quanta Red
Thermo scientific
15150

Conditions

Citation Amplex HRP pH Limit of detection
stock final unit definition stock final
mM
µM
U/ml
U/ml
BIOTEK
10
50
pyrogallol
10
0.1
7.4
4 nM (absorption 300 nm)
INVITROGEN
10
50
10
0.1
6 to 7.5
<80 nM
Towne 2004
160
0.41
7.5 to 8.5
100 nM
Zhou1997
3 ?
0.3 to 1
50 nM (10 nm optimal cond)
Mohanty1997
100
10 to 100
1
18 nM
Komary2010
1
2.5


At Oroboros Instruments we used two publications from Prof. Tretters OROBOROS MiPNet Reference Laboratory as a starting point for our development: Tretter 2012 Free Radic Biol Med and Komary 2010 Biochim Biophys Acta.

Media

Media that were designed to have some antioxidant properties will frequently consume H2O2 before it can react with Amplex (R) (Ultra)Red to form the active fluorophore resorufin. This can be shown by comparing

  • the sensitivity as determined by H2O2 additions to different media containing HRP and Amplex (R) UltraRed
  • the sensitivity as determined by addition of the actual fluorophore resorufin to the same media (signal change per added resorufin)

(Sensitivity is signal change per added H2O2 or resorufin.)

While the sensitivity measured by the addition of H2O2 drops drastically from a simple phosphate buffer to media with strong antioxidant properties, this is not observed when the sensitivity is determined by the addition of resorufin.

Since the non biological drift always observed for the Amplex /HRP system is independent of the sensitivity versus H2O2 the same (absolute) drift expressed as (calibrated) "apparent H2O2 production" will be higher in a medium with low H2O2 sensitivity as compared to a medium with high H2O2 sensitivity.

As expected some traditional KCl based media, similar to those used in ... showed a far higher sensitivity against h2O2 addition then MiR05, while some experiments indicated a lower respiration.

Media 5: Comparison MiR05Cr to MiRK03Cr

Conditions: Instrumental settings:

  • Amp.Gain 1000
  • Light intesity variable 1mA, 2mA and 5mA
  • Configuration O2K -Sensor always the same: Oxygraph AB: Sensors B-0111/B-0082 Oxygraph CD: Sensors B-0119/B-0117 Oxygraph EF: Sensors A-0004/B-0112 Oxygraph GH: Sensors A-0087/A-0088

Media: MiR05Cr and MiRK03Cr: KCl 130 mM, HEPES free acid 20 mM, KH2PO4 10 mM, MgCl2 3 mM, EGTA 0.5 mM, BSA : 0.1% pH=7 add 120 mg Creatine to 40 ml buffer

Titration:

10µl Amplex UltraRed (conc:1mM/ml -> 5µM/ml in chamber) 4µl HRP (conc:500U/ml -> 1U/ml in chamber) 2*10 µl of 56,32 µM H2O2 solution

Results/Figures: Drift and Sensitivity at least as good as MitoOx2 (=MiRK02) see Fig. 6 and 7 above


Drift vs sensitivity.png

SD drift at diff. H2O2 conc.png


no dependence on light intensity ( 1 to 5 mA) Light int. vs. calib.drift in MiR05Cr.png

Light int. vs.calib. drift in MiRK03Cr.png Fig.14 and 15



Files: can be found in P:\3D_Mario\1_development_Fluorescence\amplex\MiRK03\DatLab data 

MiRK03Cr: 1 mA: 2013-04-30 AB02/CD02/EF02/GH02 2 mA: 2013-06-18 AB01/CD01/EF01/GH01 5 mA: 2013-06-18 AB02/CD02/EF02/GH02 MiR05Cr: 1mA: 2013-04-30 AB01/CD01/EF01/GH01 2mA: 2013-06-18 AB03/CD03/EF03/GH03 5mA: 2013-07-02 AB01/CD01/ 2013-06-28 EF01/GH01 Evaluation of respirometry in MiR05Cr versus MiRK03Cr (N=2):

Fig.16 and 17 Files to be found in P:\3C_Projects_Data\2013_all data\2013_Mouse Heart\Amplex red 2013-01-22 EF02.DLD (MiR05Cr) and 2013-01-22 GH02.DLD (MiRK03Cr) 2013-01-29 EF02.DLD (MiR05Cr) and 2013-01-29 GH02.DLD (MiRK03Cr) 2013-02-05 EF02.DLD (MiR05Cr) and 2013-02-05 GH02.DLD (MiRK03Cr)_ not included in Figure

To do:? with swiched chambers with and without Creatine in low light intensity 

Fontanam 16:22, 6 August 2013 (CEST)