Comparison MiR05Cr to MiRK03Cr

Instrumental settings

  • Amp. Gain 1000
  • Light intesity variable: 1 mA, 2 mA and 5 mA
  • Configuration O2k-Fluo sensor always the same: O2k P1: Sensors B-0111/B-0082; O2k P2: Sensors B-0119/B-0117; O2k P3: Sensors A-0004/B-0112; O2k P4: Sensors A-0087/A-0088.

Media

  • MiR05Cr
  • MiRK03Cr: KCl 130 mM, HEPES free acid 20 mM, KH2PO4 10 mM, MgCl2 3 mM, EGTA 0.5 mM, BSA 0.1%, pH=7, add 120 mg Creatine to 40 ml buffer. After thawing MiRK03 a preciptate was observed, which was dissolved after 20 min shaking.

Titrations

  • 10 µl Amplex UltraRed (stock: 1 mM -> 5 µM final in chamber)
  • 4 µl HRP (stock: 500 U/ml -> 1 U/ml final in chamber)
  • 2*10 µl of 56,32 µM H2O2 solution

Results/Figures

Drift vs Sensitivity:

Drift vs sensitivity.png

Fig.6

SD drift at diff. H2O2 conc.png

Fig.7


--> Drift and sensitivity at least as good as MitoOx2 (=MiRK02)


Drift vs light intensity

Light int. vs. calib.drift in MiR05Cr.png Light int. vs.calib. drift in MiRK03Cr.png

Fig.14 and Fig. 15


--> MirK03, MiR05: drift not dependent on light intensity ( 1 to 5 mA)

compare to "MitoOx2":

Drift vs light MitoOx2.png

Evaluation of respirometry in MiR05Cr versus MiRK03Cr (N=2):

Titrations

  • 10 µl Amplex UltraRed (stock: 1 mM -> 5 µM final in chamber)
  • 4 µl HRP (stock: 500 U/ml -> 1 U/ml final in chamber)
  • 2*5 µl of 28.16 µM H2O2 solution : 0 µM, 0.07040 µM, 0.1406 µM

O2 Flux per mass MiR05Cr vs MiRK03Cr.png Flux Control Ratio MiR05Cr vs MiRK03Cr.png

Fig.16 and 17


Drift signal ratios.png

Fontanam 16:22, 6 August 2013 (CEST)