== Measuring hydrogen peroxide ==
See Measuring H2O2 production with any brand similar to  Amplex(R) REed see [[Amplex red]] and especially the accompanying Discussion page.
For an overview of methods to mneasure H2O2 concentration or H2O2 production see [[Measuring hydrogen peroxide]].


== Simultaneous measurement of TPP and fluorescence ==
== Older series ==


'''Question:''' Will it be possible to measure JO2, fluorescence AND TPP simultaneously, or just JO2/fluorescence vs JO2/TPP?
=== O2k Series A to C ===


* The Oxygraph-2k [[O2k-Main Unit#O2k-Series|Series]] A to C do not support the O2k-Fluo LED2-Module. We thank you for your understanding.


'''Answer:''' Background
The main unknown is the chemical compatibility of the methods. We still have to find out whats the effect of Amplex red and its reaction product on the TPP signal.
If this is not a big problem there are technical issues: a black TPP stopper will be required and then still light entering the chamber through the TPP electrode might be a problem. Producing the TPP electrodes in black is presumable not a very good idea because then it will no longer be possible to "see" the position of the membrane. I think the technical problems are solvable but first we have to check for chemical compatibility.


'''Update:'''
=== O2k-Fluo LED2-Module, Series A ===
First exploratory experiment:


A chamber was equipped with TPP electrodes (black stoppers) and a fluorescence module for measuring H2O2 via the Amplex Red method. A standard TPP calibration was carried out.
''' Setting the LED intensity directly at the Fluorescence-Control Unit'''
Amplex red / HRP addition resulted only in a minor signal in the TPP channel, a further TPP calibration showed no obvious large negative effects on the performance of the TPP electrode. An H2O2 calibration of the fluorescence signal was carried out: The '''sensitivity''' for H2O2 was reduced by a '''factor of 3 to 4''' as compared to  measurements without TPP electrodes. At the same time the noise  was drastically increased. In the comparative experiment (without TPP electrodes) no random noise was seen - digital resolution was the limiting factor. Based on this comparison, '''noise''' increased in the experiment with TPP electrodes at least by a '''factor of 20''', probable more.
Use the small supplied screwdriver to change the position of the switch on the front panel of the O2k-Fluo Control Unit, using the table below:
It will be possible to estimate this factor with a higher precision after running an experiment without TPP electrodes at a higher gain (higher digital resolution).


It was further tested whether the presence of TPP+ alone (without electrodes) could explain this behavior but, if there were any effects at all, they were quite small.
{| class="wikitable" style="text-align: center;"
|-
! Position
! scope="col" width="20" |0
! 1
! scope="col" width="20" |2
! scope="col" width="20" |3
! scope="col" width="20" |4
! scope="col" width="20" |5
! scope="col" width="20" |6
! scope="col" width="20" |7
! scope="col" width="20" |8
! 9
|-
! scope="row"| LED current [mA]
|| off || 0.02 || 0.5 || 1|| 2|| 5|| 10|| 20|| 30|| set by DatLab
|}


Peroxide additions corresponding to a concentration change of 110 nM (110 pmol/ml) were still well visible but not additions corresponding to a 22 nM (22 pmol/ml) concentration change.


This was only one exploratory experiment (2 chambers)!
[[Image:BB-Bioblast.jpg|left|40px|link=http://www.bioblast.at/index.php/Bioblast:About |Bioblast wiki]]
 
== Popular Bioblast page ==
Based on this preliminary data:
::: [[O2k-Fluo LED2-Module]] has been accessed more than
An important point is why one would want to measure TPP and H2O2 simultaneously in the same chamber: If this is "only" for time / sample economy the results above strongly discourage this idea. Parallel measurement e.g one chamber TPP / one chamber H2O2 are preferable.
::::* 50,000 times (2019-12-11)
If there is a very strong need to measure both parameters on the same sample (e.g. if strong in-homogeneity among samples in this respect is suspected) and a comparable high H2O2 flux is expected the simultaneous measurement might be principally doable, especially if further optimizations are possible. E.g.: Maybe with the TPP electrode in the chamber black stirrers become important again? However, this certainly would involve some methodological work and the results would be second class at the best.
::::* 45,000 times (2018-10-18)
 
::::* 40,000 times (2016-12-12)
 
::::* 30,000 times (2016-01-10)
What would be your driving interest for simultaneous measurements?
::::* 25,000 times (2015-08-20)
::::* 20,000 times (2014-12-08)
best greetings
::::* 15,000 times (2014-06-03)
Mario
::::* 10,000 times (2013-08-08)
 
::::*  5,000 times (2012-07-04)
--[[User:Fasching Mario|Fasching Mario]] 10:49, 23 February 2012 (CET)
 
 
'''Question'''
 
How many parameters can be measured at once?
 
 
'''Answer'''
 
The O2k-system is very flexible. If you take cultured cells or isolated mitochondria, split them into the two chambers of the O2k, then you can maximize the number of simultaneously measured parameters:
In both chambers oxygen concentration and oxygen flux (O2k-Core; http://www.bioblast.at/index.php/O2k-Catalogue:_O2k-Core );
in both chambers fluorescence, with either identical fluorophores and identical optical probes, or different ones for different parameters in each chamber (H2O2, ATP production, Ca2+, mt-membrane potential, with the potential to extend these possibilities; http://www.bioblast.at/index.php/O2k-Fluorescence_LED2-Module );
 
depending on fluorophore compatibility, additionally one potentiometric electrode can be inserted into each chamber (pH, TPP+, Ca2+; http://www.bioblast.at/index.php/O2k-TPP%2B_and_Ca2%2B_ISE-Module );
 
or in one chamber fluorescence and the other chamber NO (http://www.bioblast.at/index.php/O2k-NO_Amp-Module )
Importantly, oxygen concentration is not only measured, but oxygen levels can also be controlled. This will mark a new dimension in our evaluation of ROS production, with measurements spanning the entire ‘normoxic’, hyperoxic and deep hypoxic range.
 
Best wishes 
Erich
 
== Oxygraph Series A to C ==
 
*The Oxygraph-2k, [[O2k-Main Unit#O2k-Series|Series]] A to C, can currently not be equipped with an O2k-Fluorescence LED2-Module. We thank you for your understanding.
 
 
[[User:Fasching Mario|Fasching Mario]] 11:55, 25 January 2012 (CET)
 
== Other fluorophores ==
 
'''Question:'''  In the brochure, specific mention is made of Amplex Red, Mg green,  Safranin, and Ca green.  Will other fluorophores work (e.g., MitoTracker  or MitoSOX)?  Is the module tuneable at all?
 
'''Answer:''' For some application no (or little) "tuning" might be necessary, see below for your examples.
 
a.) General
 
The O2k-Fluorescence LED2-Module is "tunable" with two different levels of involvement:
The detected emission wavelength can be changed totally  by changing the filter in front of the photo diode. Similarly, some limited fine tuning of the effective excitation wavelength range is possible by changing the filters in front of the LED. Since filters (plastic films) are easily exchangeable this procedure is open to user innovation, maybe with a little help from us.
In contrast,  to drastically change the excitation wavelength  a different sensor with a different LEd has to be used. At the moment we offer [[Fluorescence-Sensor Green]] and [[Fluorescence-Sensor Blue]]. There is limited opportunity for the customer to modify this part for herself, though we might offer a third sensors for UV excitation in the future.
 
b.) Examples:
 
Remarks: a.) light intensity always has to be optimized to make sure the fluorophore considered is not affected significantly by photo bleaching. b:) Many fluorophores  are typically used for imaging applications. Be aware that the O2k-Fluorescence LED2-Module is NOT an imaging tool but quantifies fluorescence derived from the entire chamber.
 
'''MitoSOX''': According to my tables, MitoSOX  has an excitation maximum of 510 nm and an emission maximum of 772 nm. I do not have excitation or emission spectra right now to check if anything unusual or weird is going on with MitoSOX but just from the wavelengths this sounds easy.  You could start with [[Fluorescence-Sensor Blue]] plus both the LED filter and photo diode filters from [[Filter Set Saf]]. Due to the huge separation of excitation and emission it is probably possible to use no LED filter at all. It should be checked if this affects the form of any calibration curve in any way.
 
'''Mitotracker''': For me it would be interesting to learn the intended non-imaging application for Mitotracker? Also there are at least three different Mitotracker fluorophores: "green", orange", and "red". Which one do you want to use?--[[User:Fasching Mario|Fasching Mario]] 09:57, 21 May 2012 (CEST)
 
== Coupling the O2k to optical instruments: Geometry ==
 
For  various analytical techniques it is desired to couple optical sensors  and light sources to the O2k chamber. The initial approach to do so was  to insert the optical probe via a custom designed black stopper from  above directly into the chamber, see … This approach (light and  detection from above) requires also the use of black stirrer
However, in Tony Hickey's ([[Hickey 2012 J Comp Physiol B]]) and in our own  development of the fluorescence module ([[Fasching 2011 Abstract  Berlin]]  it became clear that for most applications in the visible  range it is a better and easier strategy to introduce the light via the  front window and detect the light via the same way. The Duran glass of  the O2k chamber has very high transmission down to at least 350 nm and  is probable usable even further down, see  [http://www.duran-group.com/en/about-duran/duran-properties/optical-properties-of-duran.html  Duran optical properties ]
The “via the window” approach has several important benefits:
# the optical sensor does not “see” the stirrers. Therefore, for comparable small optical sensor diameters there was no need to use black stirrers. In fact using white stirrers increased overall sensitivity without introducing any disturbances.
# no physical contact of the sample with the optical sensor necessary
# the space on top of the stopper is not taken up by the optical sensor, allowing for easier titrations and at least potentially for simultaneous use of additional electrodes introduced via the stopper
The black but otherwise regular stopper is required to shield the chamber from outside light.
Development  with this approach was facilitated by the (rather lucky) fact that the  diameter of the O2k chamber window is identical to the chamber diameter,  therefore a stopper with o-ring designed to be introduced into the  chamber from above, can also be placed in the chamber window for initial  experiments. As a permanent, final sensor a more stable solution should  be used however.
We used this configuration not only with our LED  based fluorescence module but also to couple a full spectrofluorometer  to the O2k chamber.
We think this might be worth to try also for  these groups that need a full spectrofluorometer or other light guide  based detectors / sources  attached to the oxygraph.
--[[User:Fasching Mario|Fasching Mario]] 09:26, 17 February 2012 (CET)

Latest revision as of 16:11, 11 December 2019

Previous Product ID 34000-01


Older series

O2k Series A to C

  • The Oxygraph-2k Series A to C do not support the O2k-Fluo LED2-Module. We thank you for your understanding.


O2k-Fluo LED2-Module, Series A

Setting the LED intensity directly at the Fluorescence-Control Unit Use the small supplied screwdriver to change the position of the switch on the front panel of the O2k-Fluo Control Unit, using the table below:

Position 0 1 2 3 4 5 6 7 8 9
LED current [mA] off 0.02 0.5 1 2 5 10 20 30 set by DatLab


Bioblast wiki

Popular Bioblast page

O2k-Fluo LED2-Module has been accessed more than
  • 50,000 times (2019-12-11)
  • 45,000 times (2018-10-18)
  • 40,000 times (2016-12-12)
  • 30,000 times (2016-01-10)
  • 25,000 times (2015-08-20)
  • 20,000 times (2014-12-08)
  • 15,000 times (2014-06-03)
  • 10,000 times (2013-08-08)
  • 5,000 times (2012-07-04)