== Oxygraph Series A to C ==
Previous Product ID 34000-01


'''Question:''' Why can only one chamber be used for fluoresecence measurements with O2k Series A to C?


'''Answer:''' The oxygen channel from one chamber will be used to record the fluorescence signal form the other channel. Therefore only signals from one chamber can be recorded.
== Older series ==


'''Question:''' What are the performance limitations when using the O2k-Fluorescence Module with O2k Serial A to C?
=== O2k Series A to C ===


'''Answer:'''
* The Oxygraph-2k [[O2k-Main Unit#O2k-Series|Series]] A to C do not support the O2k-Fluo LED2-Module. We thank you for your understanding.
* Only one chamber can be used for simultaneous fluorescence and respiration measurements.
 
* The amplification of the fluorescence signal is done outside the oxygraph in a special amplification box. The resulting signal stability is usually good and the same approach has been used to measure NO for O2k series A to C. However, signal stability may not reach the high level achieved when using O2k Series D and above.
 
* Performing the measurement and analyzing the data is a little bit different from the way it can be done with O2k Series D and above. The fluorescence Signal will actually be found in the signal labeled "oxygen" from the unused chamber and different amplification levels will be available. The situation is very similar to measuring NO with O2k Series A to C, see [[H MiPNet15.05 NO-Manual]].
=== O2k-Fluo LED2-Module, Series A ===
[[User:Fasching Mario|Fasching Mario]] 11:55, 25 January 2012 (CET)
 
''' Setting the LED intensity directly at the Fluorescence-Control Unit'''
Use  the small supplied screwdriver to change the position of the switch on the front panel of the O2k-Fluo Control Unit, using the table below:
 
{| class="wikitable" style="text-align: center;"
|-
! Position
! scope="col" width="20" |0
! 1
! scope="col" width="20" |2
! scope="col" width="20" |3
! scope="col" width="20" |4
! scope="col" width="20" |5
! scope="col" width="20" |6
! scope="col" width="20" |7
! scope="col" width="20" |8
! 9
|-
! scope="row"| LED current [mA]
|| off || 0.02 || 0.5 || 1|| 2|| 5|| 10|| 20|| 30|| set by DatLab
|}
 
 
[[Image:BB-Bioblast.jpg|left|40px|link=http://www.bioblast.at/index.php/Bioblast:About |Bioblast wiki]]
== Popular Bioblast page ==
::: [[O2k-Fluo LED2-Module]] has been accessed more than
::::* 50,000 times (2019-12-11)
::::* 45,000 times (2018-10-18)
::::* 40,000 times (2016-12-12)
::::* 30,000 times (2016-01-10)
::::* 25,000 times (2015-08-20)
::::* 20,000 times (2014-12-08)
::::* 15,000 times (2014-06-03)
::::* 10,000 times (2013-08-08)
::::*  5,000 times (2012-07-04)

Latest revision as of 16:11, 11 December 2019

Previous Product ID 34000-01


Older series

O2k Series A to C

  • The Oxygraph-2k Series A to C do not support the O2k-Fluo LED2-Module. We thank you for your understanding.


O2k-Fluo LED2-Module, Series A

Setting the LED intensity directly at the Fluorescence-Control Unit Use the small supplied screwdriver to change the position of the switch on the front panel of the O2k-Fluo Control Unit, using the table below:

Position 0 1 2 3 4 5 6 7 8 9
LED current [mA] off 0.02 0.5 1 2 5 10 20 30 set by DatLab


Bioblast wiki

Popular Bioblast page

O2k-Fluo LED2-Module has been accessed more than
  • 50,000 times (2019-12-11)
  • 45,000 times (2018-10-18)
  • 40,000 times (2016-12-12)
  • 30,000 times (2016-01-10)
  • 25,000 times (2015-08-20)
  • 20,000 times (2014-12-08)
  • 15,000 times (2014-06-03)
  • 10,000 times (2013-08-08)
  • 5,000 times (2012-07-04)
Cookies help us deliver our services. By using our services, you agree to our use of cookies.