Talk:O2k-chamber

From Bioblast

Previous Product ID 23100-21

Bioblast wiki

Popular Bioblast page

O2k-Chamber has been accessed more than

  • 5,000 times (2014-08-25)

Different fluxes in left and right chmaber

Question: In our experiments we get consistently higher fluxes in chamber A as compared to chamber B.

Answer: To exclude instrumental causes please follow the procedure described in the section Calibration and quality control of the OroboPOS (O2k-SOP) in MiPNet06.03 POS-Calibration-SOP.

Possible instrumental and protocol related causes for fluxes differing between chambers.

  • wrong O2 calibration in one chamber
  • wrong chamber volume calibration: The total amount of oxygen in the chamber with the smaller volume will be smaller. Therefore the O2 concentration will drop faster resulting in a higher flux.
  • wrong background parameters: wrong background parameters will cause a O2 concentration dependent offset to the flux, the value of the offset will be independent of the absolute value of the flux. E.g. a constant offset of 10 pmol /ml s at ca 100 ยตM O2 concentration and 20 pmol/ml s at 300 ยตM O2 concentration could be caused by wrong background parameters, but a difference of always 20% at very different absolute values of the flux can not be caused by wrong background parameters. In any case it is advisable to check the background parameters ( redo an instrumental O2 background experiment).
  • hydrophobic inhibitors
  • biological contamination