Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Channel

From Bioblast


high-resolution terminology - matching measurements at high-resolution


Channel

Description

The Oxygraph-2k is a MultiChannel instrument. A signal channel is defined in DatLab in terms of the recorded signal type (channel type) and the position A (left) or B (right) of the O2k-Chamber. Channel types are the oxygen channel (O2, amperometric), the additional amperometric channel (Amp), and the potentiometric channel (pX). Each of the two O2k chamber positions supports these three channel types. In addition, O2k system channels are recorded, which are the barometric pressure, block temperature, environmental temperature, an additional external temperature, and the Peltier power. In contrast to the signal channels, the system channels relate to the entire O2k.

Abbreviation: F7

Reference: MiPNet19.18A O2k-Start, MiPNet17.05 O2k-Fluo LED2-Module, MiPNet15.03 O2k-MultiSensor-ISE, MiPNet15.05 NO-Manual


MitoPedia O2k and high-resolution respirometry: O2k hardware, DatLab