|description='''Filter-Cap''': Each [[Fluorescence-Sensor Green]] and [[Fluorescence-Sensor Blue]] is equipped with a removable Filter-Cap for exchange of optical filters for the optical pathways from the LED to the sample and from the sample to the photodiode.
|description='''Filter-Cap''': O2k-Fluo LED2-Module (O2k-Series D to G) sensors ([[Fluorescence-Sensor Green]] and [[Fluorescence-Sensor Blue]]) and O2k-FluoRespirometer (O2k-Series H) sensors ([[Smart Fluo-Sensor Green]] and [[Smart Fluo-Sensor Blue]]) are equipped with a removable Filter-Cap for exchange of optical filters for the optical pathways from the LED to the sample and from the sample to the photodiode.
|product id=44310-01
|product id=44310-01
|product type=O2k, MultiSensor
|product type=O2k, MultiSensor, O2k-FluoRespirometer
|info='''''>> More details? Consult'' [[O2k-Fluorescence LED2-Module]]'''
|info='''''>> More details? Consult'' [[O2k-Fluorescence LED2-Module]]''', '''[[Smart Fluo-Sensor]]
|product image=[[Image:Filter-Cap.JPG|180px|right]]
|product image=[[Image:Filter-Cap.JPG|180px|right]]
}}
}}
Line 15: Line 15:


=== Mounting a Filter-Cap ===
=== Mounting a Filter-Cap ===
* '''Applies to''': O2k-Fluo LED2-Module (O2k-Series D to G) Fluorescence Sensor; O2k-FluoRespirometer (O2k-Series H) with Smart Fluo-Sensors
* Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements. Β 
* Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements. Β 
* Remove all filters and store them in the filter box labeled for this Filter Set. Β 
* Remove all filters and store them in the filter box labeled for this Filter Set. Β 

Revision as of 12:27, 25 July 2017

                



Filter-Cap

O2k-Catalogue

Description Filter-Cap: O2k-Fluo LED2-Module (O2k-Series D to G) sensors (Fluorescence-Sensor Green and Fluorescence-Sensor Blue) and O2k-FluoRespirometer (O2k-Series H) sensors (Smart Fluo-Sensor Green and Smart Fluo-Sensor Blue) are equipped with a removable Filter-Cap for exchange of optical filters for the optical pathways from the LED to the sample and from the sample to the photodiode.
Product ID 44310-01
Type O2k, MultiSensor, O2k-FluoRespirometer
Link >> More details? Consult O2k-Fluorescence LED2-Module, Smart Fluo-Sensor
Image
Filter-Cap.JPG

O2k-Guide

Fluorescence-Sensor lettered.jpg

Selecting a Filter Set

The Filter-Cap of each sensor can be removed for application of various filter combinations on the same optical sensor.

Mounting a Filter-Cap

  • Applies to: O2k-Fluo LED2-Module (O2k-Series D to G) Fluorescence Sensor; O2k-FluoRespirometer (O2k-Series H) with Smart Fluo-Sensors
  • Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements.
  • Remove all filters and store them in the filter box labeled for this Filter Set.
  • Insert the filters form the selected Filter Set: The round filters of each Filter Set fit to the round window of the Filter-Cap and cover the LED, the rectangular filters fit into the rectangular window of the Filter-Cap and cover the photodiode.
  • Mounting: Hold sensor and filter cap in a vertical position above you. Align the Filter-Cap with the Filter-Cap Guide (small steel rod) protruding from the sensor. Press the Filter-Cap onto the sensor without rotational movements.


O2k-Guide: Setup - next step Connect Fluorescence-Sensor to O2k