Filter-Cap
Description | Filter-Cap: Each Fluorescence-Sensor Green and Fluorescence-Sensor Cyan is equipped with a removable Filter-Cap for exchange of optical filters, which is possible independently for optical pathways from the LED and to the photodiode. |
---|---|
Product ID | 44310-01 |
Type | O2k, MultiSensor, O2k-Fluorescence LED2-Module |
Link | O2k-Fluorescence LED2-Module |
Image |
O2k-Manual
Mounting a Filter-Cap
The Filter-Cap of each sensor can be exchanged for application of different filter combinations on the same optical sensor.
- Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements. Insert a filter into either or both Filter-Cap windows for the LED or photodiode.
- Mounting: Align the Filter-Cap with the Filter-Cap Guide (small steel rod) protruding from the sensor. Press the Filter-Cap onto the sensor without rotational movements.
Setup - next step: Connect Fluorescence-Sensor to O2k