Difference between revisions of "Malate"
Line 18: | Line 18: | ||
'''Preparation of 500 mM stock solution''' (dissolved in H<sub>2</sub>O): | '''Preparation of 500 mM stock solution''' (dissolved in H<sub>2</sub>O): | ||
::1) weigh 536.5 mg of L-Malic acid | ::1) weigh 536.5 mg of L-Malic acid | ||
::2) add 3 ml H<sub>2</sub>O | ::2) add 3 ml H<sub>2</sub>O | ||
::3) neutralize with 10 N KOH (approx. 750 µl) | ::3) neutralize with 10 N KOH (approx. 750 µl) | ||
::4) adjust final volume to 5 ml and divide into 0.2 ml portions. | ::4) adjust final volume to 5 ml (in 5 mL volumetric glass flask)and divide into 0.2 ml portions. | ||
::5) store frozen at -20 °C | ::5) store frozen at -20 °C | ||
Revision as of 10:23, 25 November 2010
Description
Mitochondrial respiration depends on a continuous flow of substrates across the mitochondrial membranes into the matrix space. Glutamate and malate are anions which cannot permeate through the lipid bilayer of membranes and hence require carriers, which is also true for pyruvate. Malate alone cannot support respiration of mt-preparations.
Abbreviation: M
Reference: MiPNet11.04; MiPNet09.12
Labels:
Regulation: Substrate; Glucose; TCA Cycle"Substrate; Glucose; TCA Cycle" is not in the list (Aerobic glycolysis, ADP, ATP, ATP production, AMP, Calcium, Coupling efficiency;uncoupling, Cyt c, Flux control, Inhibitor, ...) of allowed values for the "Respiration and regulation" property.
HRR: Chemicals; Media"Chemicals; Media" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property.
Application in HRR
M: Malate (L-Malic acid, C4H605); Sigma M 1000, 100 g, store at R.T.: FW = 134.1)
Preparation of 500 mM stock solution (dissolved in H2O):
- 1) weigh 536.5 mg of L-Malic acid
- 2) add 3 ml H2O
- 3) neutralize with 10 N KOH (approx. 750 µl)
- 4) adjust final volume to 5 ml (in 5 mL volumetric glass flask)and divide into 0.2 ml portions.
- 5) store frozen at -20 °C
Oxygraph-2k Manual Titrations: MiPNet09.12
- Titration volume: 5 µl using a 25 µl syringe
- Final conc. in 2 ml O2k-chamber: 2 mM